Note:  The maximum reading the device can show is 0.40%.  A true BAC of 0.40% or above may result in 
the onset of coma or even death by respiratory arrest.  A reading of 0.40% may indicate an acutely 
dangerous condition, but may also indicate a sampling error (such as liquid alcohol blown into the 
device).  Please carefully read the 

Warning section

 of this manual before operating the device. 

 
 

- General Information about DUI/DWI - 

The legal limit for intoxication (DUI/DWI) is 

0.08%

 BAC across most of the USA.  According to the 

American Medical Association (AMA), a BAC of 

0.05%

 is enough to impair the ability to drive.  You 

should never operate a motor vehicle when you are impaired by the consumption of alcohol, drugs or 
lack of proper rest.   

 

- Warnings and Error Codes - 

Warnings and Precautions 

 

WAIT AT LEAST 20 MINUTES AFTER EATING OR DRINKING BEFORE TESTING.

 A breathalyzer 

examines “lung air” samples, not “mouth air” to calculate alcohol content. Any “contaminants” in the 
mouth, including alcohol, food, smoke, or even excess saliva, may interfere with testing. This 
includes alcohol-based mouthwash or similar products. 

This is the most common cause of improper 

use and incorrect results.

 

 

Avoid testing in high wind or enclosed spaces. Poor ventilation may also lengthen the time required 
for warm up between tests.

 

 

Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor may become 
damaged.

 

 

Do NOT use the AlcoScan AL2500 as a tool to drink and drive.

 

 

Keep out of extreme temperatures and out of the reach of children.

 

 

To maintain accuracy, the sensor inside the AlcoScan AL2500 must be recalibrated every 200 tests 
or at least once a year (whichever comes first)

. Large numbers of "high positive" (high alcohol) 

readings may require recalibration of the sensor much sooner than normally expected.  To 
recalibrate, you must mail your device to the address at the end of this manual.  Please contact us 

at 

[email protected]

 

for details and related fees.  

 

 

 

 

Содержание Alcoscan AL2500

Страница 1: ...1 DIGITAL PERSONAL BREATHALYZER AlcoScan AL2500 AlcoScan AL2500 Operator s Manual...

Страница 2: ...2...

Страница 3: ...quired The AlcoScan AL2500 is intended to measure alcohol in human breath Readings obtained by this device are used in the diagnosis of alcohol intoxication The accuracy of this device has been establ...

Страница 4: ...f a sample is not taken within 20 seconds you will hear four 4 beeps and the unit will shut itself off You can also press the POWER button again to manually shut off the unit 3 Analysis and Display On...

Страница 5: ...air samples not mouth air to calculate alcohol content Any contaminants in the mouth including alcohol food smoke or even excess saliva may interfere with testing This includes alcohol based mouthwash...

Страница 6: ...ne batteries included Warm up Time Less than 1 minute Testing Capacity Approx 200 with new batteries DOT Approved YES Operating Temperature 50 104 degrees Fahrenheit or 10 40 degrees Celsius Warm Up L...

Страница 7: ...cohol concentration depends on many variables including environmental factors such as air quality wind humidity temperature etc and health conditions of the test subject A low g dL or BAC reading does...

Страница 8: ...RY INCLUDING BUT NOT LIMTED TO LOSS OF USE LOSS OF REVENUE OR INCOME LOSS OF ACTUAL OR ANTICIPATED PROFITS INFLUDING LOSS OF PROFITS FROM A CONTRACT LOSS OF THE USE OF MONEY LOSS OF ANTICIPATED SAVING...

Отзывы: