background image

CAM-ENG-60-000521-A 

© 2013

Содержание B-2 Series

Страница 1: ...Calibration manual V3 SERIES B 2 Alcohol gas standard ...

Страница 2: ...s International Inc and are used under license Alcohol Countermeasure Systems is the trading style of Alcohol Countermeasure Systems Corp 2013 Alcohol Countermeasure Systems The information disclosed in this document is the valuable property of Alcohol Countermeasure Systems and all copyright and other proprietary rights to this document are reserved No reproduction of this document is permitted w...

Страница 3: ...and should be avoided Examine the canister and valve for any damage and the expiration date on the label Observe all cautions and safety warnings found on the canister Never remove or alter canister labels Never modify the delivery tube in any way Always remove the valve and install the protective cap on cylinders when not in use Store cylinders in a cool well ventilated area away from sources of ...

Страница 4: ...lled the date and time cannot be changed 6 In the Select method menu press Next to scroll to Manual and press Select 7 In the Select type menu press Next to scroll to Dry Gas and press Select The Select value menu is displayed Skip the next section and read Calibrating a handset continued B BY MAINTENANCE CODE NO DOWNLOAD STATION NOTE The handset must be connected to the ECU 1 Power the handset on...

Страница 5: ...ns the gas standard concentration options Verify your canister label press Next to scroll to 130 PPM and press Select This enters the Altitude menu Altitude is preset to 0 meters 2 Determine the altitude of your exact location NOTE Find exact altitude online Altitude may vary within general region or city always use exact value 3 If your altitude is 100 m or lower leave at 0 m Press Accept 101 m o...

Страница 6: ... the calibration is complete If 3 or 4 tests are required a Calibration OK message is displayed when calibration is complete 8 Do I or II I For calibration with download station disconnect the handset from the download station and connect back to the ECU II For calibration without download station press the right button to finish there is also the option to retry by pressing the left button CALIBR...

Страница 7: ...ed in the Select Value menu The alcohol gas standard is expired All connections to and from the gas valve are not secured properly The value entered in Altitude menu is incorrect ALTITUDE REFERENCE TABLE The altitude adjustment is in increments of 200 meters Round the altitude up or down to the closest value as follows IF ALTITUDE IS THEN SET TO 0 100 m 0 101 300 m 200 301 500 m 400 501 700 m 600 ...

Страница 8: ...CAM ENG 60 000521 A 2013 ...

Отзывы: