ALCOLOCK V3 SERIES B-2
2
CALIBRATING A HANDSET
Calibration is possible both with a download station or by maintenance code (without
a download station).
Do
A
or
B
.
A. WITH A DOWNLOAD STATION
1.
Connect the download station to a power source.
2.
Disconnect the handset from the ECU.
3.
Connect the download station to the handset with a coiled cable.
The handset automatically powers on.
4.
Press the
left
button on the handset to select
Cal
.
The current handset date and time are displayed.
5.
In the
Date / Time
screen, do
I
or
II
:
I.
If the date and time are correct, press
OK
– the
Select method
menu is displayed.
II.
If the date or time is incorrect, press
Set
to automatically synchronize handset time
with download station time, and press
OK
– the
Select method
menu is displayed.
NOTE: If a tachograph is installed, the date and time cannot be changed.
6.
In the
Select method
menu, press
Next
to scroll to
Manual
and press
Select
.
7.
In the
Select type
menu, press
Next
to scroll to
Dry Gas
and press
Select
. The
Select
value
menu is displayed.
Skip the next section and read: “Calibrating a handset (continued).”
B. BY MAINTENANCE CODE (NO DOWNLOAD STATION)
NOTE: The handset must be connected to the ECU.
1.
Power the handset on by holding down either the
right
or
left
navigation button.
2.
Access the menu screen by pressing and holding the
left
button.
NOTE: Where available, press and hold the
right
navigation button to exit a current
screen / menu.
3.
Press
Next
to scroll to
System Maintenance
, and press
Select
.
4.
Input the daily System Maintenance Code (obtained from ITE). To do so:
•
Press
[+]
repeatedly to increment a digit.
•
Press
Next
to move the cursor to the next digit.
5.
With the code entered, press
Next
to move the cursor to
OK
, and press
Select
.
6.
In the service menu, press
Next
to scroll to
Calibration
, and press
Select
.