3.3. Warm Up

 

The Determinator must warm to operating temperature before a sample is requested. Bringing a unit indoors in 
a cold climate will reduce the required warm up time. The message displayed while heating shows 0% to 
100% completion. 

 

 
  3.4 Camera (Optional) 

Some State Laws require the use of Cameras with some Breath Alcohol Ignition Interlock Devices.  This camera is contained within the Hand Held 
portion of your device.  State Law states that any attempt to block or obscure the camera will be considered a violation, and can be considered 
grounds for extension, and or cancelation of your Driving Privilege.  Attempts to obscure the camera will result in a failed attempt to start, delaying 
you from starting your vehicle.  With that in mind be cautious of how you hold the device while testing, be sure to hold it level with the ground, 
camera up (any other way will the device will not accept your sample) do not place your hands around the device, DO cup it or hold it on the sides, 
so as not to interfere with the built-in camera.   

 
           

How to properly hold the device during test!                                 Holding this way may result in violations!  

              

                                        

 

 
 

3.5  Provide Sample: Blow/Suck/Blow (BSB)

 

The Determinator requests a breath sample at turn-on and at random intervals while you drive. 

  Do not provide a sample until requested 

  Two beeps are sounded for attention 

before “Please Provide Sample” is seen/heard. 

  Breathe in deeply to fill lungs. 

  Provide the sample in three steps: 

o

  STEP #1: Blow: short, sharp, pulse of air until you hear a beep. 

o

  STEP #2: Suck: short, sharp, pulse of air until you hear a beep. 

o

  STEP #3: Blow: long, continuous stream of air to empty the lungs. Keep the pressure above/below 

the min/max set points. Three tones may be heard: 

  Low tone : blow harder 

  Medium tone : try to maintain this 

  High tone : blow softer 

  When the tone stops, stop blowing, sample is complete. Wait for the voice prompt to start the vehicle. DO 

NOT attempt to start the vehicle until instructed to do so, attempting a start before being prompted may result in a violation. 

 
 

Breath Sample Procedure:

 

 

 

3.5.1Sample Errors

 

If an error is detected while taking a sample, the 

Determinator says “INVALID SAMPLE” 

at the same time the display will scroll one of 

the following messages to report the actual error and requests the user retry the sample. Possible errors include: 

  "Err: Pressure Not Stable" => Prior to Step #1 - do not blow into the unit until requested 

  "Err: Suck Not In Time" => Step #2 (short suck) sample not provided in allotted time 

  "Err: Blow Not In Time" => Step #3 (long blow) sample not provided in allotted time, or you may be blowing too hard for it to register 

  "Err: Low Sample Pressure" => Step #3 - pressure was too low during final breath sample 

  "Err: High Sample Pressure" => Step #3 - pressure was too high during final breath sample 

  "Err: Sample Temp" => Step #3 

–temperature out of range, Do not leave the device in direct sunlight as it may overheat the device 

  "Err: Fuel Cell Stability" => At the start of a sample, the alcohol measurement device was not ready 

Содержание ADS Determinator

Страница 1: ...ond the service due at midnight notification Using that buffer may cause your required time on the BAIID to be extended There is a difference between the device stating SERVICE DUE and LOCKDUE At every start up the device will display the number of days till service service due XX days pay attention to this number if you have an event or other issue which results in an early return for service tha...

Страница 2: ... This will allow you to get familiar with the retest procedure without the distraction of driving Once you become familiar with the procedure of the retest you will easily be able to perform it while driving but until you get familiar give yourself the chance to learn it safely Remember you have 3 minutes to provide the sample never do anything while driving which will distract you from your prima...

Страница 3: ...a breath sample Display messages and voice prompts are provided by the Hand Held to instruct a user 2 2 Determinator Vehicle module The Vehicle module is wired to the vehicle s electrical system It is out of sight of the user and provides the connection to the vehicle The coiled cord is permanently attached to the Vehicle module and connects tothe Hand Held 2 3 Determinator Mouthpieces Along with ...

Страница 4: ...ile youdrive Do not provide a sample until requested Two beeps are sounded for attention before Please Provide Sample isseen heard Breathe in deeply to fill lungs Provide the sample in three steps o STEP 1 Blow short sharp pulse of air until you hear a beep o STEP 2 Suck short sharp pulse of air until you hear a beep o STEP 3 Blow long continuous stream of air to empty the lungs Keep the pressure ...

Страница 5: ...apply 7 Failing Tests Tampering Circumvention Early Service If you fail a start test due to the presence of alcohol then the device will enter a temporary time out 3 10 min when this occurs you should immediately determine the reason correct it and retest after an appropriate time or the temporary time out has passed A person tampering with circumventing or otherwise misusing this Ignition Interlo...

Страница 6: ...nator s menu and steps through options Also the 2 key Menu Button NO Exit the Determinator menu Also the 3 key DEMOEXIT press YES to exit Demo Mode SERV DUE how many days until required to return to ServiceCenter VOLUME press YES to increase the voice volume values 1 10 TIME HMS display unit s time DATE YMD display unit s date DAYS2CAL number of days until a Service Center calibration is required ...

Страница 7: ...t was in possession of the repair center If you are doing the work yourself keep any receipts for any items you purchase to complete the work Q The State has sent me a letter asking for a request of explanation for anevent s A This will occur if you do something that is considered a violation of state law We cannot tell you how to answer these requests for explanation you will need to refer to you...

Страница 8: ...CTE The unit has an optional time restriction entered Unit will only operate between certain hours Does not apply to Kansas DISABLED AMBIENTTEMP System error the internal ambient temperature is out of expected range Turn off ignition roll windows down if hot and retry DISABLED HEATING VOLTAGE System error the heaters encountered an unexpected condition this generally may be an indicator ofa dead o...

Отзывы: