6/6
C 71
Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual
Basic internal calibration of the leak
detector
Edition 03 - December 01
GB 00969 -
When the auto-calibration is complete, the unit
is ready to start a cycle.
Note: it is possible to start a test cycle after an auto-
calibration failure.
F1
F3
F2
F4
READY FOR CYCLE
INLET ...
STDBY
F1
F3
F2
F4
PLEASE,OPEN He LEAK
OK -> press NEXT
F1
F3
F2
F4
ON TEST PORT!
OK -> press NEXT
These 2 screens appear
alternately
6
End of the
Close the valve of the calibrated leak (if there
process
is one).
Validate the operation.
Note: If these operations are not been done in the minute
which follows the appearance of these screens, the auto-
calibration is automatically stopped. A message informs the
operator of this stop.
NEXT