Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual
Basic internal calibration of the leak
detector
Edition 03 - December 01
GB 00969 -
Calibration
Connect the external calibrated leak.
preparation
4
F1
F3
F2
F4
PLEASE,OPEN He LEAK
LEAK VALUE:
1.0E-07
EXT
Θ
(°C/°F):20/68
OK -> press NEXT
F1
F3
F2
F4
ON TEST PORT!
Leak value: 1.0E-07
EXT
Θ
(°C/°F):20/68
OK -> press NEXT
These 2 screens appear
alternately
Note: If the operator press
,all the parameter of the
calibrated leak programmed display.
Open the valve of the calibrated leak (if there
is one).
F2
F1
F3
F2
F4
PLEASE,OPEN He LEAK
LEAK VALUE:
1.0E-07
EXT
Θ
(°C/°F):20/68
OK -> press NEXT
F3
Set the ambiant temperature.
Adjust a value
C 20
F1
F3
F2
F4
PLEASE,OPEN He LEAK
LEAK VALUE:
1.0E-07
EXT Q (°C/°F):20/68
OK -> press NEXT
F3
5
Validate the auto-calibration process
Different screens appear during the process
giving internal parameters values.
NEXT