The skin should be free of oils, grease, gels 
or other debris, to prevent reactions with the 
skin or the structure of the material.

• 

The orthosis should fit firmly but not 
too tight, so as not to restrict the blood 
circulation and adversely affect nerve and 
lymph vessels. Excessive compression is 
therefore to be avoided.

• 

Combination with other products is currently 
not provided for or is to be agreed with the 
manufacturer in writing.

• 

The brace is not intended for single use, 
but is intended for multiple use by a single 
person.

• 

The product as delivered is not sterile.

• 

Contact your physician immediately in the 
event of an allergic reaction.

• 

Please note that cushioned sections can heat 
up under direct sunlight. Protect the orthosis 
from direct sunlight if necessary.

• 

Currently there is no test for flammability. 
Exercise caution when using the orthosis 
in the direct vicinity of open flames such as 
lighters and cigarettes.

1.11. Warranty

We assume the warranty for the brace for a 
period of 3 months. The brace is a medical 
rehabilitation device and must not be used 
for any purpose other than the intended, as 
described in the Instructions for Use. Changes 
to the brace or other applications require the 
written consent of the albrecht GmbH. If this 
is not obtained, the manufacturer may not 
honor the guarantee. If you are using single 
joints or other components, these should be 
used as intended. If changes or modifications 
(e.g. additional mounting holes) are made 
to the individual parts or components, the 
manufacturer‘s warranty no longer applies. 
The removal of or damage to the QM seal will 
also void the warranty.

1.12. Transfer of the brace

The brace is not intended for single use, but 
rather is intended for multiple use by a single 
person. We do not recommend transfer to 
other users. Should this be desired however, 
please ensure to pass on the care and cleaning 
instructions and have the brace checked by an 
authorized specialist dealer for safe and proper 
operation.

1.13. Disposal

The brace contains recyclable materials 
without toxic or other harmful substances or 
other environmentally hazardous substances. 
Provided it is not contaminated with infectious 
germs, the brace can be deposited in the 
normal waste disposal. To be sure, consult 
your specialist orthopedics dealer.

EN

User Instructions

www.albrechtgmbh.com

5

Содержание 782A-1

Страница 1: ...U S E R I N S T R U C T I O N M K S O S T E O S P I N A L O R T H O S I S ...

Страница 2: ... Features 4 1 8 Indications 4 1 9 Contra indications 4 1 10 Safety Instructions 4 1 11 Warranty 5 1 12 Transfer of the brace 5 1 13 Disposal 5 2 Individual adjustments 6 3 Cleaning maintenance and disinfection 7 3 1 Care 7 4 Technical data 8 5 Choice of size 8 6 Size chart and article numbers 8 MKS Osteo User Instructions 00800 03032900 INTERNATIONALE SERVICE HOTLINE 2 ...

Страница 3: ...ble please contact the appropriate specialist retailer or albrecht GmbH directly before you use the brace This particularly applies to the safety instructions 1 3 Mode of operation The MKS Osteo Spinal orthosis functions according to the reclination principle The two shoulder straps attached to the back plate straighten the vertebral column to relieve the thoracic spine The wide lumbar bandage str...

Страница 4: ...ain caused by poor posture e g rounded shoulders Poor posture Hollow round back Chronic lumbago caused by hyperlordosis Spondylodiscitis Facet syndrome Degenerative changes in the lumbar spine Tumor Condition after intervertebral disc surgery Muscle insufficiency Abdominal muscle insufficiency Spondylitis For all other indications a physician must be consulted 1 9 Contra indications The brace is o...

Страница 5: ...e used for any purpose other than the intended as described in the Instructions for Use Changes to the brace or other applications require the written consent of the albrecht GmbH If this is not obtained the manufacturer may not honor the guarantee If you are using single joints or other components these should be used as intended If changes or modifications e g additional mounting holes are made ...

Страница 6: ...nsure reliable effectiveness Always make sure that the MKS Osteo is a smooth and tight fit However in order to prevent constrictions along the edges it shouldn t be over tightened 1 Take the ready for use MKS Osteo previously adjusted by your orthopaedist out of the packaging and put it on like a backpack 2 Close the body belt using the hand loops Make sure that the belt fits correctly i e flat an...

Страница 7: ... machine washable The MKS Osteo can be dried in the open air In the case of more severe soiling a replacement set of textile parts is available Wipe down surfaces with a cloth soaked with disinfectant Wet completely and do not wipe off Spray inaccessible surfaces When spraying ensure complete wetting A mild alcohol based disinfectant is recommended Ask your physician or pharmacist when selecting a...

Страница 8: ...B 1 2 70 80 cm 18 cm 782C 2 782A 2 782B 2 3 80 90 cm 18 cm 782C 3 782A 3 782B 3 4 90 105 cm 18 cm 782C 4 782A 4 782B 4 5 105 120 cm 18 cm 782C 5 782A 5 782B 5 6 120 135 cm 18 cm 782C 6 782A 6 782B 6 MKS Osteo reduced height Size Body circumference Abdominal height Height of back piece XS 43 cm S 46 cm M 50 cm L 56 cm Art No Art No Art No Art No 1 60 70 cm 14 cm 782D 1N 782C 1N 782A 1N 782B 1N 2 70...

Страница 9: ...u are required to immediately report any serious incident involving the use of this medical device to the manufacturer and the responsible authorities Please find our contact details on the back of this brochure User Instructions www albrechtgmbh com 9 ...

Страница 10: ...MKS Osteo User Instructions 00800 03032900 INTERNATIONALE SERVICE HOTLINE 10 ...

Страница 11: ...User Instructions www albrechtgmbh com 11 ...

Страница 12: ...ERSION EN 08 2022 albrecht GmbH CHIEMSEESTRASSE 81 D 83233 BERNAU AM CHIEMSEE T 49 0 8051 96129 0 F 49 0 8051 96129 30 INFO ALBRECHTGMBH COM WWW ALBRECHTGMBH COM Medical device Manufacturer albrecht GmbH ...

Отзывы: