Instruction manual
- 11 -
during the 10 minutes.If the leak on the screen is not
greater 0.13,it will show OK. Or it will show failure.
5.5.3.4 "Tp" i
s the temperature difference between the max.
temperature and the min. temperature during the
cycle. If it is greater 3
℃
it will show void. That means
the result is void. You need run the test again after the chamber has cooled down.
6. Advanced Setting
6.1 Enter the setting menu
6.1.1
Hold START button and then turn on the main power
sw
itch. Keep holding for about5 seconds, you will enter
setting menu.
6.1.2 Press PROGRAM button, you can select to set S1,S2,S3,
S4. The chosen option will blink. Press START button and enter the sub menu.
6.2 S1 setting
You can adjust units of pressure and temperature as you
like, also of the date and time.
6.2.1
For example,you can adjust temperature unit as
following:Press PROGRAM button to switch setting option,when the temperature
unit is blinking,press TEMP button,and it will change between F and ℃
. Press
PROGRAM button again to switch to the next setting option.
6.2.2
The other option can be adjusted as the same way above.
6.2.3
After finish it, press START button to exit this menu, all of the data you have
Содержание STE16L-C
Страница 26: ...Instruction manual 23 APPENDIX 1 Characteristics of the feeding water ...
Страница 29: ...Instruction manual 26 ...
Страница 30: ...Instruction manual 27 ...
Страница 31: ...Instruction manual 28 ...
Страница 32: ...Instruction manual 29 APPENDIX 4 ELECTRICAL DRAWING ...
Страница 33: ...Instruction manual 30 APPENDIX 5 ...
Страница 34: ...Instruction manual 31 HYDRAULIC DRAWING ...