AirTest TR-5200 Скачать руководство пользователя страница 4

WARRANTY:

ATI AirTest Technologies, Inc ( the company) warrants all  instruments manufactured
by them to be free from defects in materials and workmanship under normal use and
service  for  a  period  of  one  year.  Electrochemical  and  Catalytic  Bead  sensors
purchased from sensor manufacturers will be warranted to the extent offered by the
manufacturer. The company’s obligation of this warranty shall be limited to the repair
or replacement of any part found to be defective. The company shall not be obligated
to repair or replace units that are found to be in need of repair because of damage,
(other than normal wear and tear), tampering or modification. The detector must be
returned to AirTest suitably packaged and with freight prepaid. Liability is limited to
the  replacement  cost  of  a  new  unit.  In  no  case  shall  the  company  be  liable  for  any
consequential or incidental damages however caused. No agent, dealer or employee
of the company has the authority to increase or alter the obligations or limitations of
this warranty.

SPECIFICATIONS:

Sensing Element

Catalytic Bead

Standard Range

0 to 100 % LEL
Combustible gas

Gas Sampling Method

Diffusion

Linearity

Linear

Operating Temperature

-20

°

C to +50

°

C

-4

°

F to +122

°

F

Dimensions(LxWxD) in. TR5200-6.2 x 4 x 2.4  TR5500-6.8 x 3.5 x 4

     cm.                 15.6 x 10 x 6

      17.3 x 9 x 7.6

Humidity ( non-cond.)

0 to 90%

Sensor Life Expected

3+ years

Calibration Interval

12 months

Warm-up Time

< 3 minutes

Response Time

< 1 minute

Power Requirement

12 to 30 VDC

Power Consumption

< 2 watts

Output

4-20 mA

Enclosure Material  

TR5200-PVC, TR5500 Cast Aluminum

Weight

TR5200-16 ounces, TR5500-4 lbs.

Approvals

TR5200-CSA/NRTL
TR5500-CSA/UL Cl.1, Div.1

6

ATI AirTest Technologies, Inc

#9, 1520 Cliveden Avenue

Delta, BC, V3M 6J8

Ph. 604-517-3888
Fx. 604-517-3900

e-mail [email protected]
website www.airtesttechnologies.com

Содержание TR-5200

Страница 1: ...BRATION The sensor transmitter must be connected to the controller Allow the sensor three to five minutes to warm up before calibrating 1 Connect negative lead of voltmeter to test point C YELLOW Connect positive lead of voltmeter to test point D WHITE Apply zero gas to the sensor ambient air may be used if it is certain there are no contaminants in the area Adjust VR2 ZERO to 0 0Volts 0 020V 2 Co...

Страница 2: ...o high For the TR 5200 this is 4 25 volts For the TR5500 it is 3 50 volts THE SIX TEST POINTS ON THE LEL BOARD ARE AS FOLLOWS A GROUND BLACK C ZERO TEST POINT YELLOW D SPAN TEST POINT WHITE E 4 20mA TO VOLTS RED B SENSOR VOLTAGE F SYSTEM 10 VDC 0 5V B C D SIG VR1 SENSOR V SEN VR2 VR3 ZERO SPAN A F E VR4 CURRENT A1 INPUT CONTROLLER A1 Fig 1 SENSOR V CONCENTRATION OF CALIBRATION GAS 100 MAXIMUM RANG...

Страница 3: ...A input Controller Note Supply voltage must be 10 volts higher than maximum controller voltage On a 0 to 10 volt controller the minimum supply voltage required is 20 volts For Controller with voltage input R Value 5 volt input use 250 ohm 10 volt input use 500 ohm External Load resistor TR 5200 WIRING TR5200WWire tcw dcm B C D SIG VR1 SENSOR V SEN VR2 VR3 ZERO SPAN A F E VR4 CURRENT A1 INPUT B C D...

Страница 4: ... company be liable for any consequential or incidental damages however caused No agent dealer or employee of the company has the authority to increase or alter the obligations or limitations of this warranty SPECIFICATIONS Sensing Element Catalytic Bead Standard Range 0 to 100 LEL Combustible gas Gas Sampling Method Diffusion Linearity Linear Operating Temperature 20 C to 50 C 4 F to 122 F Dimensi...

Отзывы: