CALIBRATION (Con’t

)

NOTE: The system voltage, measured between point A 

(BLACK)

 (negative) & point F

(RED)

 (positive),  is factory set.  Do not change this voltage value.

The sensor voltage, measured between point A 

(BLACK)

 (negative) & point B

(YELLOW)

 (positive), is factory set and the value is sensor type dependent. Do

not change this voltage value. Damage to the sensor will result if the voltage
is set too high. For the TR-5200 this is 4.25 volts.  For the TR5500 it is 3.50
volts.

               

THE SIX TEST POINTS

ON THE LEL BOARD

ARE AS FOLLOWS:

A - GROUND                            BLACK

C - ZERO TEST POINT            YELLOW

D - SPAN TEST POINT            WHITE

E - 4-20mA TO VOLTS              RED

B- SENSOR VOLTAGE 

F - SYSTEM 10 VDC +/- 0.5V    

B

C

D

+ -

SIG

VR1

SENSOR V

SEN

VR2 VR3

ZERO SPAN

A

F

E

VR4

CURRENT

+ -

A1

INPUT

CONTROLLER

+

-

A1

Fig. 1

SENSOR

            

       

V = 

% CONCENTRATION OF CALIBRATION GAS

100 % (MAXIMUM RANGE)

[

]

X 1.6 + 0.4

2

LAYOUT – TR5200

MODEL TR-5200

4.80

4.00

2.40

4.00

6.60

AirTest Technologies Inc

Drawn By

Drawing No.

Rev

TR-5200 LAYOUT

TR5200WLayout.tcw

dcm

COMBUSTIBLE

GAS

DETECTOR

3

Содержание TR-5200

Страница 1: ...BRATION The sensor transmitter must be connected to the controller Allow the sensor three to five minutes to warm up before calibrating 1 Connect negative lead of voltmeter to test point C YELLOW Connect positive lead of voltmeter to test point D WHITE Apply zero gas to the sensor ambient air may be used if it is certain there are no contaminants in the area Adjust VR2 ZERO to 0 0Volts 0 020V 2 Co...

Страница 2: ...o high For the TR 5200 this is 4 25 volts For the TR5500 it is 3 50 volts THE SIX TEST POINTS ON THE LEL BOARD ARE AS FOLLOWS A GROUND BLACK C ZERO TEST POINT YELLOW D SPAN TEST POINT WHITE E 4 20mA TO VOLTS RED B SENSOR VOLTAGE F SYSTEM 10 VDC 0 5V B C D SIG VR1 SENSOR V SEN VR2 VR3 ZERO SPAN A F E VR4 CURRENT A1 INPUT CONTROLLER A1 Fig 1 SENSOR V CONCENTRATION OF CALIBRATION GAS 100 MAXIMUM RANG...

Страница 3: ...A input Controller Note Supply voltage must be 10 volts higher than maximum controller voltage On a 0 to 10 volt controller the minimum supply voltage required is 20 volts For Controller with voltage input R Value 5 volt input use 250 ohm 10 volt input use 500 ohm External Load resistor TR 5200 WIRING TR5200WWire tcw dcm B C D SIG VR1 SENSOR V SEN VR2 VR3 ZERO SPAN A F E VR4 CURRENT A1 INPUT B C D...

Страница 4: ... company be liable for any consequential or incidental damages however caused No agent dealer or employee of the company has the authority to increase or alter the obligations or limitations of this warranty SPECIFICATIONS Sensing Element Catalytic Bead Standard Range 0 to 100 LEL Combustible gas Gas Sampling Method Diffusion Linearity Linear Operating Temperature 20 C to 50 C 4 F to 122 F Dimensi...

Отзывы: