3.0  SAFETY PRECAUTIONS

READ SAFETY PRECAUTION BEFORE INSTALLATION

Incorrect installation can cause serious damage or injury

WARNING

CAUTION

Failure to observe a warning may result in death 
or serious injury.

The appliance must be installed in accordance 
with applicable codes and regulations.

Failure to observe a caution may result in injury 
or equipment damage.

NOTE: 

Failure to observe notes may result in equipment 

or property damage.

This device MUST be installed by qualified personnel in 
accordance with the manufacturer’s installation instructions 
and local authority requirements. All electrical work must be 
undertaken by a licensed electrician.

   

WARNING

• 

Before installing or servicing this unit, turn all power 
OFF and have unit unplugged. Electrical shock can 
cause personal injury or death. 

• 

Do not exceed voltage rating. Severe damage to device 
and / or fire could result. Verify proper voltage before 
use.

• 

Never expose eyes or skin to UV-C light from any 
source. Looking directly at the UV-C light may cause 
permanent eye damage or blindness. 

• 

Never operate the unit out of the duct or plenum. The 
system is designed for a closed duct system. 

• 

Do not mount near a supply or return air opening.

• 

All duct openings in direct line of sight of the UV light 
must be sealed with aluminium tape or equivalent 
sealing methods.

• 

UV light will break down plastic materials not rated for 
UV exposure e.g. wire insulation, flexible duct, drain 
pans and humidifiers. 

 

o

The unit must be mounted at least 76mm from 
non-UV rated materials.

 

o

If it is necessary to mount the unit in direct line 
of sight of these types of materials, they must be 
shielded with aluminium foil, aluminium foil tape, 
or sheet metal.

• 

The UV-C lamp contains mercury. Never touch a 
broken lamp or the internal contents of the lamp. 

• 

After opening the carton, unpack and inspect the unit 
and lamp for damage. Do not install this unit if any 

damage is noticeable. If the lamp is damaged, cracked 
or broken, do not touch lamp or other contents in 
carton. Close carton and return to place of purchase 
immediately.

• 

In case of accidental exposure to internal contents of 
lamp, wash hands and any other affected skin area with 
mild detergent. Remove affected clothing immediately. 
If skin surface is pierced, seek medical attention 
immediately.

• 

Do not touch the lamp with bare hands. If you touch the 
glass of the lamp, you should wipe the glass clean with 
isopropyl alcohol and a lint free wipe. If the lamp is not 
cleaned, the film of hand oils causes a reduction in the 
light intensity.

• 

To prevent water damage or electric shock, do not 
mount unit under a humidifier, or near the condensate 
drain pan.

• 

DO NOT mount this unit outdoors. This product is 
designed for indoor installation only.

• 

For use on gas fired, oil fired, electric and split system 
heat pump forced air systems. This product is NOT 
designed to be mounted outside on gas fired or heat 
pump packaged units.

 

Содержание Uvhealth BIO2000

Страница 1: ...INSTALLATION MANUAL GERMICIDAL UV C AIR PURIFICATION ...

Страница 2: ... WARNING Disclaimer This product is not sold as a medical device and is not for the purpose of diagnosis of any disease or condition nor for use in the mitigation treatment or prevention of any disease or condition Copyright 2021 AirSmart Pty Ltd www uvhealth com au TABLE OF CONTENTS 1 0 PRODUCT DESCRIPTION 5 2 0 PARTS LIST 5 3 0 SAFETY PRECAUTIONS 6 4 0 SIZING AND SELECTING UNITS 7 5 0 UVHEALTH B...

Страница 3: ...tion Unlike room air purifiers UVHealth is concealed in your roof space or ductwork and provides continuous whole of building air purification BIODEFENCE SYSTEM FOR YOUR HOME OR WORKPLACE UVHealth units will improve system performance and on going maintenance by reducing Organic build up on HVAC R Cooling Heating Coils by illumination with UV C Germicidal Rays ...

Страница 4: ...ce of 76mm away from flexible ductwork and other non UV suitable materials Transition box BIO1000 or BIO2000 Alternate installation location for additional units Multiple units can be installed to increase the coverage area Cleaner healthier air Kills germs Removes odours Retrofit to most ducted air condiotioners Whole of home or office solution Improves heat exchanger performance by preventing bi...

Страница 5: ...ycomb cell technology removes VOCs and odours Carbon adsorption layer traps VOCs and odours The UVHealth BIO2000 uses high intensity ultraviolet lamps to produce germicidal UV C band light rays This UV C light sterilises and reduces airborne microorganisms and volatile organic compounds VOC s by a combination of photocatalytic chemical reaction and carbon absorption as air passes through your air ...

Страница 6: ...ill break down plastic materials not rated for UV exposure e g wire insulation flexible duct drain pans and humidifiers o The unit must be mounted at least 76mm from non UV rated materials o If it is necessary to mount the unit in direct line of sight of these types of materials they must be shielded with aluminium foil aluminium foil tape or sheet metal The UV C lamp contains mercury Never touch ...

Страница 7: ...acity L s 750 1000 1500 1750 2000 2250 2500 2750 3000 No of BIO1000 req d 1 0 2 1 0 3 2 1 0 plus No of BIO2000 req d 0 1 0 1 2 0 1 2 3 Approx coverage m2 150 200 300 350 400 450 500 550 600 Figure 3 UVHealth BIO1000 and BIO2000 size selection chart Figure 4 UVHealth BIO1000 and BIO2000 application selection chart Contaminant Type Power Supply Bacteria Mould Viruses VOCs Odours 24V DC 120 230VAC BI...

Страница 8: ...orced air ducted systems DO NOT install outdoors Product is not designed to be exposed to outdoor elements 1 The ballast is factory set to 120 Volt AC Make sure that the voltage is set to 230 Volt AC by sliding the voltage selector switch in Figure 5 to the 230 Volt position The plug must be replaced and wiring must be done to applicable electrical codes Figure 5 Location of the voltage selector s...

Страница 9: ... 460 460 300 Figure 10 Minimum dimensions for transition piece The unit will reduce organic build up on an HVAC cooling heating coil by illumination with germicidal UV C rays Locate the unit such that the maximum surface area of the coil is illuminated while installing as close to the coil as possible without damage to the coil Use care to avoid damage to the coil and any wiring or other component...

Страница 10: ...unting Plate to the opening by folding the inner flaps of the plate into the opening and fully back against the inside of the opening in the ductboard either from the sides or from top and bottom of the plate Finish the Ductboard Mounting Plate installation by sealing around the edges of the plate to the ductboard using mastic foil tape or other material suitable for sealing duct joints e Insert t...

Страница 11: ...VHealth BIO2000 back into the duct mounting plate and tighten the mounting screws 7 Replace the front electrical cover and plug the quick connect power cord back into the unit mounting screws 6 2 REPLACING THE LAMP It is recommended the lamp is replaced after 12 months of continuous service To maintain maximum output intensity the lamp should be replaced annually Replace only with a genuine replac...

Страница 12: ...neycomb or carbon media with the specified replacement parts only When sliding the replacement pieces back into the unit check and make sure the honeycomb is placed on the inside while the carbon media is touching the outside side panel 8 Once the replacements are slid all the way into the opening attach the end bracket with previously removed hex screws 9 Carefully insert the lamp through the gro...

Страница 13: ...tion Value UV C Germicidal Lamp 50 watts UV C Length in Duct 451 mm UVHealth BIO2000 PCO TiO2 Semiconductor 2 26 m2 Weight 1 63 kg Adjustable Voltage 120 or 230 volts Lamp Life Efficiency Output Up to 2 years Average Static Pressure Drop 4 98 pa Application air flow range for the UVHealth BIO2000 up to 1000 L s 17 6 kW 9 0 REPLACEMENT PARTS LIST Replacement Part Number Part Description BIO2000PCO ...