[Read the Important Safety Rules section before operating this equipment.]
18
MN165-1 Rev G
6.0 PRODUCT SPECIFICATIONS
VisionAire
™
2
and
VisionAire
™
3
Flow Specification
s
(1)
½ lpm – 3 lpm
⅛
lpm – 2 lpm
±10% of indicated setting, or 200 mL, whichever is greater
(1)
Oxygen Concentration
(1)
90% (+ 5.5% / - 3%)
Electrical Requirements
115 VAC / 60 Hz, 3.0 A
230 VAC / 50 Hz, 1.5 A
230 VAC/ 60 Hz, 1.5 A
Power Consumption
VisionAire 2 = 200 Watts (+/- 10%)
VisionAire 3 = 175 Watts (+/- 10%)
Sound
40 dba
Dimensions
14.1 in. W x 11.5 in. D x 20.8 in. H
(35.8 cm W x 29.2 cm D x 52.8 cm H)
Weight
30 lbs (13.6 kg)
Electrical Shock
Protection
Class II, Type B
Environmental Limit
Conditions
(2)
Operating conditions:
5
o
C to 40
o
C (41
o
F to 105
o
F ) at
altitudes up to 10,000 ft.(523mmHg) above sea level.
Storage:
-20
o
C - 60
o
C (-4
°
F - 140
°
F)
Relative Humidity
: up to 95% RH
non - condensing
(1)
Performance based at 21
°
C (70°F) at nominal operating pressure range: 0-7 kPa
(0-1 psig backpressure).
Maximum outlet pressure : 55 kPa. ( 7.98 psig)
(2)
Operating outside of these operational specifications can limit the concentrators ability
to meet Oxygen Concentration specification at higher liter flow rates.