RECHARGEABLE BATTERY INFORMATION
PLEASE READ CAREFULLY BEFORE USING THE BATTERY.
1. You should fully charge your battery for at least 8 hours before using
it for the first time. Protect the battery from high temperatures when
charging, operating and storing.
2. For optimal charging efficiency, the battery should be charged at an
ambient temperature of 10~30°C. If battery is charged outside of the
recommended temperature range, you may notice erratic functioning
of the charging indicator light.
3. If the battery is used at an ambient temperature over 45°C, the
nebulizer may shut down automatically to prevent overheating.
If this occurs, allow the device and battery to cool down.
4. When the nebulizer power indicator light turns to orange (indicating
low power), detach the battery from the nebulizer until charging.
5. Only charge and recharge this battery with the AC adaptor supplied
with your nebulizer. DO NOT attempt to use any other device as it
may damage your battery and will void its warranty.
6. If the battery is not to be used for a long period of time, it should be
recharged every three (3) months to maintain optimal performance.
ABOUT THE TWO-VALVE FEATURE
The mouthpiece included with this unit
features a two-valve system. During expiration,
the expiration valve will open, allowing the
patient's exhaled air to pass to atmosphere,
avoiding contamination of the device. During
inspiration, the valve will close, maximizing
the amount of inhalation and reducing any
waste of drug to a minimum.