CLEANING INSTRUCTIONS
®
In order to ensure the proper functioning of your MedPro AirPort™
Portable Nebulizer System, it should be thoroughly cleaned regularly.
AFTER EVERY INHALATION
1. Turn off the compressor by pressing the ON / OFF button. Remove
the battery block or unplug the power cord from the power outlet
and then from the unit.
2. Remove the mouthpiece (or mask, mask adapter, air tubing
assembly), nebulizer cap and micronization cone.
3. Rotate and take out the nebulizer chamber and clean all mobile
parts under running water.
4. Wipe the unit’s body by using a clean, moistened soft cloth.
IMPORTANT: Never clean with benzene, thinner or any other
flammable chemical.
5. Wash all other parts of the nebulizer system, especially the parts
that have been in contact with the nebulized drug (including the
mouthpiece, mask, mask adapter and/or air tubing assembly),
nebulizer, micronization cone and nebulizer chamber under
running water.
6. Rinse thoroughly all mobile parts (excluding the compressor unit
and battery pack) under lukewarm running water (max. 60 degrees)
for 10 minutes and wipe with a clean dry cloth.
7. Let air-dry completely and store in proper conditions until next use.
PERIODICALLY, TO REDUCE RISK OF INFECTION
Repeat all the procedures mentioned above. The mouthpiece, mask
adapter, complete nebulizer chamber assembly can be sterilized by
keeping in boiling water for 10 minutes. Do not boil the complete
device or any other parts.