If the device does not operate when plugged in and the MAIN POWER SWITCH
is in the “ON” position, check the fuse on the rear panel of the device. Unplug the
device and remove the fuse from its housing, or contact your local authorized
dealer for further information.
Important: When replacing a blown fuse, to protect against fire hazard replace the
fuse with one of the identical type and rating (2.5 AMP T250VACH SLO-BLO). If the
fuse blows again, return the pump to your dealer for servicing.
CAUTION
: There are no parts inside the device that are user-
serviceable. There is an electrical shock hazard if the pump assembly
is disassembled. Refer all service to qualified personnel.
CAUTION
: Keep away from environment of CT or MRI.
CAUTION
: Keep away from explosive or flammable anesthetic gas.
Troubleshooting
Error Messages/Alarms
High Pressure: This can occur if there is a kink or other obstruction in the tubing
connecting the garments, or if the connector port is blocked. The LCD screen will
read E01, the red LED light will illuminate, and an alarm will sound.
Low Pressure: This can occur if the garment is not properly secured to the
connector port or there is a hole in the garment causing a leak. The LCD screen will
read E02, the red LED light will illuminate, and an alarm will sound.
Software System Error: This can occur if an internal software system error occurs
causing the device to be inoperable. The LCD screen will read E03, the red LED
light will illuminate, and an alarm will sound.
NOTE
: If you receive any of these error messages, press the
START|STOP
BUTTON
, which will begin deflation of the garments. Then, contact your local
authorized dealer or AIROS Medical.