background image

 
Page  17 of  23 

USER MANUAL 

ADS intelligent 

 

 

 

 

Softwareversion: 

0.95 

 
 
 
 

 

9.9

 

Movement history 

 

 

 
 
The  movement  history  recorded  movement  data  will  be  recorded  day  by  day  and  shown  in 
two  rows,  an  upper  row  from  midnight  till  12  o'clock,  and  in  one  lower  row  from  12  to  24 
o'clock.  The operator  can select  the  date/day  by pressing  the buttons  plus  (+) and minus  (-). 
 
The  current  position  of  what  is  shown  in  the  display  and  in  the  schematic  mattress  is  marked 
by a small  red bar above the motion  monitor  graph  (1).  
 
By  touching  the  time  graph  the  operator  can  scroll  horizontally  from  "early"  to  "late"  into  one 
day.  The  yellow  line  (3)  indicates  the  necessary  pressure  level.  During  the  scrolling,  the 
schematic  mattress  in  the  display  also  changes  the  cells,  according  to  the  pressure  inside  the 
cell.   
 
Remark:  the  pink  area  indicates,  that  during  this  time  change  your  settings  in  the  system  by 
the  operator has been done.  This  time  should  not  be used for  analyses  of movement.   
 

Содержание ADS Intelligent

Страница 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Страница 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Страница 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Страница 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Страница 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Страница 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Страница 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Страница 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Страница 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Страница 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Страница 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Страница 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Страница 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Страница 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Страница 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Страница 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Страница 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Страница 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Страница 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Страница 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Страница 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Страница 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Страница 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Отзывы: