11

PROGRAMMING MENUS

BLOWER CONTROL –

 This menu is preset at the factory to meet O.S.H.A. face velocity

 

    standards while in AUTOMATIC mode. From this screen you can:

 

    • Turn the blower on/off.

 

    • Switch from automatic to manual blower speed based on the application.

 

    • Change the output of the blower.

TIMER – 

   This menu allows the operator to turn the timer on/off and to program an

 

    appropriate event time. Once the operator has programmed the desired time,

 

    the Timer can be activated by touching the Timer Quick Key on the front of the

 

    controller.

TIME/DATE –

The operator can change displayed time and date.

ALARM SETTINGS –

All alarm points for face velocity, HEPA fi lter and gas detection have

 

    been preset by the manufacturer and should not be altered.

ALARM SCREEN –

  Allows the operator to turn the low airfl ow, HEPA, and gas alarms on/off

 

    in this screen. This controls both the audible and the visible alarms. If the

 

    operator turns off the alarm in this screen, nothing (audible or visible) will alert

 

    you if the airfl ow is below the selected face velocity, the HEPA fi lter is clogged,

 

    or the bonded carbon fi lter fails.

  

   

NOTE:

 The operator can deactivate the audible alarm, when it sounds, by

 

    touching the alarm button on the controller. The visible alarm will still work.

FILTER –  

This menu shows the serial number and the purchase date of the fi lter. For

 

    every new fi lter installed, a new fi lter serial number must be entered. Refer to

 

    the “MAINTENANCE” section for complete instructions. The fi rst fi lter serial 

 

    number has already been entered for you prior to shipment.

BLOWER HOURS

 – Displays the hours and minutes that the blower has been active.

GAS DETECTOR

 – (for carbon enclosures only) A solid state hydrocarbon detector warns

 

    the user of fi lter breakthrough. Alarms are audible and visible. 

ADDITIONAL INFORMATION

    

•  Keep the right hand side of the unit away from the wall and 

    

   turbulent places. Never install enclosures near doorways or high-traffi c areas.

    

•  To avoid confusion, each alarm has a distinctive sound and will fl ash what  

  

   the problem is on the display screen.

    

•  A few minutes are needed for the airfl ow sensor to warm up, stabilize, and  

  

   settle at the correct airfl ow speed.

       

•  Unless the operator turns the blower off (by Quick key or in the blower menu)  

  

   before turning the power off, the blower will automatically come on each time  

  

   the operator activates the unit.

    

•  Anything the operator changes within the program is automatically saved.

    

•  Allow each enclosure to run fi ve minutes, before use, to stabilize 

    

   the environment.

Содержание POWDERSAFE 760C

Страница 1: ...6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 2005 AirClean Systems Inc ACOM03600PS C OPERATOR S MANUAL POWDERSAFE DUCTLESS...

Страница 2: ...d Particulate and Alarm Messages 7 Operating Instructions for Fumes and Vapors and Alarm Messages 8 Air ow Path 9 Controller Instructions 10 Programming Menus 10 11 Quick Keys 12 Maintenance Pre lters...

Страница 3: ...st AirClean Systems has eliminated this problem with breakthrough bonded carbon technology Bonded carbon lters are solid blocks of activated coconut shell carbon held in a solid and predictable matrix...

Страница 4: ...18 x 3 Requirements 2 3 4 Secondary HEPA Filter3 Type ACFHEPA ACFHEPA ACFHEPA Dimensions 12 x 24 x 3 12 x 24 x 3 12 x 24 x 3 Requirements 2 3 4 Bonded Carbon Filter4 ilter4 ilter Optional Type Applica...

Страница 5: ...3 Requirements 2 3 4 Secondary HEPA Filter3 Type ACFHEPA ACFHEPA ACFHEPA Dimensions 12 x 24 x 3 12 x 24 x 3 12 x 24 x 3 Requirements 2 3 4 Bonded Carbon Filter4 ilter4 ilter Optional Type Application...

Страница 6: ...Place the enclosure on the cart or the counter C Remove the pre lter for inspection before use Refer to the MAINTENANCE section of this manual for complete instructions D Connect the supplied power co...

Страница 7: ...particulate inside your enclosure B When working within the enclosure keep the front sash down and perform your procedure in the center to rear of the enclosure C If the low air ow alarm sounds and or...

Страница 8: ...substances please call our of ce to verify compatibility before use in this case your lter may have another model number Direct Dial 919 255 3220 or 800 849 0472 Monday Friday 8 00 5 00 PM EST 2 Conne...

Страница 9: ...rostatic pre ltration and HEPA ltration 3 Particulate free air moves upward to C where it enters the HEPASafe ltration area This ltration is only challenged with particulate during primary lter change...

Страница 10: ...m armed POWER Turns on off power to the controller and its keys This DOES NOT cut off power to the entire enclosure The main power switch is located above the power inlet on the back right hand side o...

Страница 11: ...ible alarm when it sounds by touching the alarm button on the controller The visible alarm will still work FILTER This menu shows the serial number and the purchase date of the lter For every new lter...

Страница 12: ...tive key for enclosure LIGHT One touch to enable uorescent light if installed TIMER User de nable lab event timer AUX Inactive key for enclosure Figure 6 AirSafe automatic safety controller layout ont...

Страница 13: ...your unit The pre lter and pre lter grid are located at the back of the unit 2 Look through the slats on the pre lter grid If the pre lter is discolored the original color is green replacement is need...

Страница 14: ...x nuts and remove the lter hold down 6 Place the spent HEPA lter into the second poly bag Wipe any residue off of the lter hold down frame and any other obvious areas Place the second rag into the bag...

Страница 15: ...cart is completely assembled 3 Repeat steps above for holes 2 and 3 4 For standard setup complete assembly using hole 4 For seated use complete assembly using hole 5 5 Tighten all bolts completely 6...

Страница 16: ...Cross Piece 2 Locking Caster 2 Standard Caster 4 Washer 8 1 Washer 4 1 x Bolt 8 2 x Bolt Standard Setup For Seated Use Mounting Holes Hole 3 Hole 5 Hole 4 Hole 1 Hole 2 ASSEMBLED CART CONFIGURATIONS...

Страница 17: ...orescent Fixture 30w 1 10 50428 Fuse 6 3a 250V Slo Blo 1 1 1 06 50481 Gas Strut 60 2 2 06 51087 Gas Strut 90 2 06 50012 Hinge Door Positioning 6 8 12 06 50053 Hinge Medium Black 3 4 5 11 52756 Keypad...

Страница 18: ...orescent Fixture 30w 1 10 50428 Fuse 6 3a 250V Slo Blo 1 1 1 06 50481 Gas Strut 60 2 2 06 51087 Gas Strut 90 2 06 50012 Hinge Door Positioning 6 8 12 06 50053 Hinge Medium Black 3 4 5 11 52756 Keypad...

Страница 19: ...tems or the dealer who sold y ou the product If the cause is determined to be a manufacturing fault the dealer or AirClean Systems will repair or replace all defective parts to restore the unit to ope...

Страница 20: ...eturn any materials without authorization from AirClean Systems Systems Products returned without prior authorization will not be accepted AirClean Systems Systems and its dealers are not responsible...

Страница 21: ...04 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 Visit AirClean Syste Syste ms on the Internet at http www aircleansystems com Email AirClean Systems at Systems at info aircleansystems co...

Страница 22: ...gh NC 27604 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 E mail info aircleansystems com Product designs are subject to change without notice 2005 AirClean Systems Inc Systems Inc ACOM03...

Отзывы: