11
PROGRAMMING MENUS
BLOWER CONTROL –
This menu is preset at the factory to meet O.S.H.A. face velocity
standards while in AUTOMATIC mode. From this screen you can:
• Turn the blower on/off.
• Switch from automatic to manual blower speed based on the application.
• Change the output of the blower.
TIMER –
This menu allows the operator to turn the timer on/off and to program an
appropriate event time. Once the operator has programmed the desired time,
the Timer can be activated by touching the Timer Quick Key on the front of the
controller.
TIME/DATE –
The operator can change displayed time and date.
ALARM SETTINGS –
All alarm points for face velocity, HEPA fi lter and gas detection have
been preset by the manufacturer and should not be altered.
ALARM SCREEN –
Allows the operator to turn the low airfl ow, HEPA, and gas alarms on/off
in this screen. This controls both the audible and the visible alarms. If the
operator turns off the alarm in this screen, nothing (audible or visible) will alert
you if the airfl ow is below the selected face velocity, the HEPA fi lter is clogged,
or the bonded carbon fi lter fails.
NOTE:
The operator can deactivate the audible alarm, when it sounds, by
touching the alarm button on the controller. The visible alarm will still work.
FILTER –
This menu shows the serial number and the purchase date of the fi lter. For
every new fi lter installed, a new fi lter serial number must be entered. Refer to
the “MAINTENANCE” section for complete instructions. The fi rst fi lter serial
number has already been entered for you prior to shipment.
BLOWER HOURS
– Displays the hours and minutes that the blower has been active.
GAS DETECTOR
– (for carbon enclosures only) A solid state hydrocarbon detector warns
the user of fi lter breakthrough. Alarms are audible and visible.
ADDITIONAL INFORMATION
• Keep the right hand side of the unit away from the wall and
turbulent places. Never install enclosures near doorways or high-traffi c areas.
• To avoid confusion, each alarm has a distinctive sound and will fl ash what
the problem is on the display screen.
• A few minutes are needed for the airfl ow sensor to warm up, stabilize, and
settle at the correct airfl ow speed.
• Unless the operator turns the blower off (by Quick key or in the blower menu)
before turning the power off, the blower will automatically come on each time
the operator activates the unit.
• Anything the operator changes within the program is automatically saved.
• Allow each enclosure to run fi ve minutes, before use, to stabilize
the environment.