SOLR
®
100% OXYGEN MASK LEVEL 3
TM0063-2
60410X
5-29
33.
Once the microphone wire is properly set, use the T8 torx and torque wrench to secure the microphone
wire retainers to the regulator to 6-inch-lbs.
Figure 5-71 Torque Microphone Wire Retainers to 6-Inch-lbs
34.
Align the two notches on the left side of the mask to the two indents on the regulator assembly.
Figure 5-72 Align the Notches on the Mask and the Regulator
35.
Insert the regulator assembly into the
left
side of the mask.
36.
Place locking washer onto regulator assembly with the raised edge facing down.
Figure 5-73 Place Locking Washer with Raised Edge Facing Down onto
Regulator
Содержание SOLR 60410X
Страница 4: ...TM0063 2 SOLR 100 OXYGEN MASKLEVEL 3 II 60410X INTENTIONALLY BLANK...
Страница 6: ...TM0063 2 SOLR 100 OXYGEN MASKLEVEL 3 IV 60410X INTENTIONALLY BLANK...
Страница 8: ...TM0063 2 SOLR 100 OXYGEN MASKLEVEL 3 VI 60410X INTENTIONALLY BLANK...
Страница 20: ...TM0063 2 SOLR 100 OXYGEN MASK LEVEL 3 5 12 60410X INTENTIONALLY BLANK...
Страница 42: ...TM0063 2 SOLR 100 OXYGEN MASK LEVEL 3 6 4 60410X INTENTIONALLY BLANK...
Страница 48: ...TM0063 2 SOLR 100 OXYGEN MASK LEVEL 3 6 10 60410X INTENTIONALLY BLANK...
Страница 49: ...SOLR 100 OXYGEN MASK LEVEL 3 TM0063 2 60410X 6 11 Figure 6 2 100 Regulator Assembly Illustrated Parts Breakdown...
Страница 51: ...SOLR 100 OXYGEN MASK LEVEL 3 TM0063 2 60410X 6 13 Figure 6 3 Exhaust Valve Assembly Illustrated Parts Breakdown...
Страница 54: ...TM0063 2 SOLR 100 OXYGEN MASK LEVEL 3 6 16 60410X INTENTIONALLY BLANK...
Страница 66: ...TM0063 2 SOLR 100 OXYGEN MASK LEVEL 3 B 60410X INTENTIONALLY BLANK...