TM0063-2
SOLR
®
100% OXYGEN MASK LEVEL 3
5-28
60410X
30.
Place the sealing ring on the regulator body.
Figure 5-68 Place Sealing Ring on the Regulator
31.
Place a dust cap on the quick disconnect.
Figure 5-69 Place Dust Cap on Quick Disconnect
32.
Using the T8 torx driver, connect the microphone cable to the regulator. Adjust the microphone wire
until the female end of the microphone wire is even with the end of the quick disconnect, ± 1/8-inch.
Figure 5-70 Ensure Microphone Wire is Even with the Quick Disconnect