DRS25 User’s Guide Chapter 6: Software Overview
0025.00.902
6-31
Nozzle Heater Temperature, Flow Rate and Temperature:
The nozzle heater temperature
(TempºC)
(B)
can be programmed from 0ºC to 450ºC.
The nozzle flow rate
(Rate%)
(A)
can be programmed from 15% to 110% (100% represents 2.75 scfm).
A flow rate range between 15% and 19% allows a maximum temperature of 325ºC.
Air / Nitrogen Option:
The pressure source connection for the upper heater can be
Air
or
Nitrogen
(D)
. Selecting the
Air
option
automatically connects to the upper heater to the main air source (large blue hose). The main air source
is also used by the vacuum and cooling systems. Selecting the
Nitrogen
option automatically connects
the upper heater to the secondary pressure connection (smaller blue hose or yellow/black hose). This
secondary pressure source is generally connected to nitrogen. If nitrogen is not available, this required
connection must be made to air.
Note:
The secondary pressure connection typically provides a more stable flow rate environment and
should be considered for all placement/soldering events.
Note:
If nitrogen conservation is a critical facilities issue, applications can be programmed to only use
the nitrogen option during process steps where stabile flow rate is essential.
(B)
(A)
(D)
Содержание DRS25
Страница 1: ...DRS25 ...
Страница 2: ......
Страница 4: ......
Страница 6: ...DRS25 User s Manual Table of Contents 0025 00 902 ...
Страница 8: ...DRS25 User s Guide Chapter 1 Introduction Getting Started 0025 00 902 1 2 ...
Страница 20: ...DRS25 User s Guide Chapter 2 Installation 0025 00 902 2 2 ...
Страница 36: ...DRS25 User s Guide Chapter 2 Installation 0025 00 902 2 18 ...
Страница 37: ......
Страница 46: ...DRS25 User s Guide Chapter 3 Machine Function Verifications 0025 00 902 3 2 ...
Страница 66: ...DRS25 User s Guide Chapter 4 Process Development for Engineers 0025 00 902 4 2 ...
Страница 78: ...DRS25 User s Guide Chapter 4 Process Development for Engineers 0025 00 902 4 14 The Cool Down Event 45 seconds ...
Страница 94: ...DRS25 User s Guide Chapter 4 Process Development for Engineers 0025 00 902 4 30 ...
Страница 96: ...DRS25 User s Guide Chapter 5 Operator Training 0025 00 902 5 2 ...
Страница 140: ...DRS25 User s Guide Chapter 6 Software Overview 0025 00 902 6 2 ...
Страница 145: ...DRS25 User s Guide Chapter 6 Software Overview 0025 00 902 6 7 6 1 Options Menu Functions View of the Options Menu ...
Страница 156: ...DRS25 User s Guide Chapter 6 Software Overview 0025 00 902 6 18 6 2 Control Menu Functions View of the Control Menu ...
Страница 214: ...DRS25 User s Guide Chapter 7 Troubleshooting 0025 00 902 7 2 ...
Страница 223: ...DRS25 User s Guide Chapter 8 Schematics 0025 00 902 8 1 8 Schematics ...
Страница 224: ...DRS25 User s Guide Chapter 8 Schematics 0025 00 902 8 2 ...
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ...DRS25 User s Guide Chapter 9 Maintenance 0025 00 902 9 2 ...
Страница 263: ...DRS25 User s Guide Chapter 9 Maintenance 0025 00 902 9 23 9 8 Wire Diagrams and Assembly Drawings ...
Страница 264: ...DRS25 User s Guide Chapter 9 Maintenance 0025 00 902 9 24 ...