The following table shows the standard values for the sensor exposure time of a child patient
DC radiator, 7 mA
Tube length 8 in. (20 cm)
DC radiator, 6 mA
Tube length 12 in. (30 cm)
60 kV
µGy
70 kV
µGy
Incisor
004 s
51
007 s
58
Pre-molar
006 s
81
010 s
82
Molar
008 s
108
013 s
107
Bite wing
009 s
121
014 s
115
• Check and adjust X-ray appliances in accordance with the following standard equipment-specific values
Pre-programmed exposure times for films with class E sensitivity on an adult patient
DC radiator, tube length
8 in. (20 cm)
DC radiator, tube length
12 in. (30 cm)
7 mA
60 kV
6 mA
70 kV
7 m A
60 kV
6 mA
70 kV
Incisor
016 s
008 s
032 s
016 s
Pre-molar
020 s
010 s
040 s
020 s
Molar
025 s
012 s
050 s
025 s
Bite wing
032 s
016 s
064 s
032 s
Pre-programmed exposure times for films with class E sensitivity on a child patient
DC radiator, tube length
8 in. (20 cm)
DC radiator, tube length
12 in. (30 cm)
7 mA
60 kV
6 mA
70 kV
7 mA
60 kV
6 mA
70 kV
Incisor
010 s
005 s
020 s
010 s
Pre-molar
012 s
006 s
025 s
012 s
Molar
016 s
008 s
032 s
016 s
Bite wing
020 s
010 s
040 s
020 s
• Check and adjust X-ray appliances in accordance with the following standard equipment-specific values
• Press selection key image plate/sensor
• Press selection key tooth symbol
• Press selection key Adult/Child patient
• Press selection key imaging parameters
– kV, and
to select other values
• Press selection key imaging parameters
– mA, and
to select other values
• Press selection key imaging parameters
– secs, and
to select other values
• Press key Save for 2 secs
The individual settings have now been saved and can be read on the display
To reset the unit to the factory settings please contact your Service
Technician