SECTION 4: REPLACEMENT PARTS
Problem
Possible Cause
Corrective Measures
7 Image not
reproduced
squarely onto
duplicating film.
7a. Original radiographs not
placed correctly onto du-
plicating film.
7a. Use care when placing
original radiographs. Make
sure the duplicating film and
radiograph are positioned
within a template grid.
7b Duplicator Door closed before
securing Retainer Plate.
7b Always close and lock
Retainer Plate before
closing Duplicator Door.
7c Retainer Plate closed too
quickly.
7c Close Retainer Plate slowly
so as not to scatter films.
8. Completely
clear films.
8a. Incorrect film placement.
8a Place duplicating film first
and then place radiograph
onto duplicating film.
8b Exposure time too long
8b Decrease exposure time
using adjustment screw
located at right rear of
Daylight Loader. (Turn screw
counter-clockwise to reduce
exposure time.)