Air Techniques Peri-Pro Скачать руководство пользователя страница 6

Page 5

Air Techniques, Inc. 

SECTION 2: MAINTENANCE

Always unplug the Duplicator from the electrical outlet prior to performing any main-

tenance. It is recommended that you remove the cover from the Peri-Pro to facilitate 

maintenance.

Bulb Replacement

1.  Open Daylight Loader view glass.
2.  Open Duplicator Door. This will expose the light source prism.
3.  Remove the 2 Phillips head screws at the top of the prism adjacent to the rubber 

bumpers.

4.  Carefully remove prism. Some flexing of the prism will  be necessary, however, do not 

bend excessively as this may cause breakage.

5.  Unscrew burnt out bulbs and replace with part number I105AI, replacement  bulbs.
6.  Replace prism and secure with the 2 Phillips head screws removed in step 3.·

Duplicator Door Removal

1.  Open Duplicator Door.
2.  Apply gentle pressure at bottom of Duplicator Door towards the left until the left hinge 

may be lifted away from its socket.

3.  Push Duplicator Door toward the right after the left hinge is freed from its hinge socket 

to remove the right hinge.

4.  Lift Duplicator Door out of Film Duplicator/Daylight Loader Assembly.

Printed Circuit Board and Reflector  Assembly Replacement.

1.  Remove Duplicator Door and remove prism as above. The PC Board and Reflector 

Assembly is now loose.

2.  Carefully pull the PC Board and Reflector Assembly towards the left while gently push-

ing on potentiometer dial on outside  of unit. The  Assembly may now be lifted.

3.  Unscrew  Phillips head screw on reflector while holding nut that secures the ground 

cable.

4.  Unplug  spade  connectors from  terminals and  remove  PC  Board  and Reflector 

Assembly.

5.  Install new PC Board and Reflector Assembly (PN92740) and

Содержание Peri-Pro

Страница 1: ...Peri Pro Intra Oral Film Duplicator Daylight Loader Operating Instructions...

Страница 2: ...aneously Has an automatically timed exposure Does not require a darkroom Duplicates size 0 I or 2 intraoral radiographs onto size 2 duplicating film Before using the Film Duplicator Daylight Loader be...

Страница 3: ...are individually wrapped and can be handled in room light during this step Some duplicating films are not individually wrapped and are m a common pouch 2 Close view glass 3 Insert hands through cuffs...

Страница 4: ...ction of the grid with the lighter side of film emulsion facing up NOTE The duplicating film has a projecting pin dot on the emulsion side Make sure this projecting dot faces up 6 On each duplicating...

Страница 5: ...hin the Daylight Loader 10 Remove the duplicating film from the Duplicating Grid and process through the Peri Pro in the usual manner 7 Close Duplicator Door and push firmly with both hands to activat...

Страница 6: ...3 Duplicator Door Removal 1 Open Duplicator Door 2 Apply gentle pressure at bottom of Duplicator Door towards the left until the left hinge may be lifted away from its socket 3 Push Duplicator Door to...

Страница 7: ...latches fully engage when closing duplicator door Wait for light source to shut off prior to opening duplicator door 2b Emulsion of duplicat ing films is placed face down 2b Always place duplicating...

Страница 8: ...hers are okay 5a One or two bulbs burnt out 5a Replace bulb See instructions page 5 5b Some of the original radio graphs are much darker than others 5b Check technique factors when x ray exposure is m...

Страница 9: ...7c Retainer Plate closed too quickly 7c Close Retainer Plate slowly so as not to scatter films 8 Completely clear films 8a Incorrect film placement 8a Place duplicating film first and then place radi...

Страница 10: ...Page 9 Air Techniques Inc NOTES...

Страница 11: ...ator and manufacturer of dental products Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support Whether the need is di...

Отзывы: