Model H-2 for VacStar 20

Congratulations! Your new HydroMiser, water recycling sys-
tem patented design incorporates many unique benefits:

•  HydroMiser reduces VacStar’s water use by up to 75%.

•  A built-in air/water separator keeps air and gases from 

entering your sanitary waste lines.

•  A built-in self-clearing drain device prevents solids 

from accumulating within the liquid storage reservoir.

•  A unique eductor fluid control system assures proper 

water flow for optimum VacStar performance.

The HydroMiser, is available as Model H-2 for the VacStar 20 
and Model H-4 used with VacStar 40.

Warning:

  The following steps must be performed 

while wearing skin and eye protection 

designed for handling typical Haz-Mat 

material. Use care at all times to prevent 

spillage.

Installation Specifications

1.  Items A, D. & E are supplied. 
2.  Item B, 2" vent to outside is not 

supplied. 

3.  Item C, 1-1/2" sanitary waste line is 

not supplied.

4.  If local codes prohibit a direct 

waste connection as shown, use 
hose supplied to route waste from 
"E" to a floor sink or to an open 
vertical 1-1/2" PVC waste line. 

Important

•  Install HydroMiser to the left, right, 

or above of VacStar.

•  Do not install HydroMiser below 

VacStar.

•  Maximum vertical distance between 

the VacStar exhaust outlet and the 
HydroMiser inlet is 3 feet. (36 in.)