AS12, AS22, AS50, AS70,
I n t e r n a t i o n a l
SITE REQUIREMENTS
Service Clearance:
Allow 12" on all sides for all models .
Ambient Temperature:
Must not exceed 105°F (40°C)
Air System Plumbing Connection:
3/8" F .N .P .T . Shut-off valve and a 6 ft . pressure hose (supplied)
Air distribution piping for all models - 1/2", type "L" or type "K" copper
If pipe volume is too great, more than 235 in
3
(3 .85L) or more than 100 ft . (30 .5m)of
1/2" diameter pipe, a pressure regulator should be installed between the main tank
and the distribution piping . Set pressure to pressure switch cut in value (factory set
at 85 PSI) .
Environmental:
Operating
Indoor use at altitudes up to 6562 ft (2000m) . Temperature 41 to 105°F (5 to 40°C) .
Supply voltage fluctuation of +/- 10% of nominal voltage .
Storage and Transport:
Temperature, 0 to 150°F (-18 to 65°C) .
Relative Humidity, 0 to 90% .
IEC 60601-1 Classification:
Protection against electric shock (6 .2):
Class I
Applied Parts (5 .9 .1, 8 .3):
There are no Applied Parts .
Protection against harmful ingress of water (6 .3):
Ordinary, IPXO
Degree of safety in the presence of flammable
anesthetics mixture with air or with oxygen or with
nitrous oxide (11 .4, 11 .5):
Not suitable .
Mode of operation (6 .6):
- All Modes Except AS100:
Continuous
- AS100 Only:
50% Duty Cycle; Maximum
Continuous ON Time of 20 Min