Page 28
13.3.2
Internal AC power fuse
The AC supply transformer on the Main PCB is protected against an internal fault by a non-
resettable thermal fuse. To avoid the thermal fuse being accidentally tripped by connecting to a
230V supply when the instrument set for 115V operation, a standard 500mA (T) 250V fuse is fitted
in position FS3 on the main PCB. Should this fuse need replacing, following such an event, remove
the case upper and replace the fuse following the instructions and diagrams of
Section 3.
13.4
Cleaning
If the meter requires cleaning use a cloth that is only lightly dampened with water or a mild
detergent. Polish the display window with a soft dry cloth.
WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE METER, NEVER ALLOW
WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE OR DISPLAY
WINDOW NEVER CLEAN WITH SOLVENTS.
13.5
Firmware Updates
The instrument firmware can be updated in the field through the USB port. It requires a PC software
utility provided by the manufacturer, and uses a HID driver that will already be installed on the PC. If
that is the only USB functionality required, download the package containing the firmware update
together with the PC utility from the manufacturer, and follow the instructions included.
Note:
The latest revisions of this manual, device drivers and software tools can be
downloaded from:
http://www.aimtti.com/support
.