Troubleshooting

For professional 

in vitro

 diagnostic use only.

•  For all professional healthcare settings
•  Place the meter into the docking/

charging station when it is not in use

•  Clean with 70% isopropyl alcohol, 10% 

dilution of commercial grade sodium 
hypochlorite or commercial surface 
decontamination preparations

•  Do not spray liquid directly to the 

meter or immerse into liquid

Nova StatStrip Glucose and ß-Ketone 

Control Solution

•  Store control material vial tightly closed under 30°C
•  Do not refrigerate or freeze.
•  Open vial stability is up to 3 months  

or until expiration date.

•  Store strips at 

15°C–30°C.

•  Use strips within 

expiration day

 or 

glucose test 

within 180 days 

of opening 

the vial and ß-Ketone strips within  

90 days

 of opening the vial.

Sample material

•  Wash patient’s hands or clean with alcohol pad and dry  

completely before taking the sample.

•  The test strip must be filled in a one go. Do not add more sample.
•  For capillary finger stick, venous whole blood and arterial samples.
•  Suitable also for neonatal samples and patients receiving intensive 

medical therapy.

•  Suitable anticoagulants: sodium, lithium, and ammonium heparin.

GLU

Nova StatStrip Glucose and ß-Ketone 

Hospital Meter

Nova StatStrip Glucose and ß-Ketone 

strips

Transfer failed: 

The meter was removed before  

data transfer was complete. Please re-dock  
the meter.

Flow error:

 The specimen was incorrectly drawn 

into the test strip due to either insufficient or incor-
rect sample application. Repeat the test with a new 
strip. If the error code persists, perform test using  
an alternate method.

Analysis canceled: 

The test has been canceled, 

repeat the test with a new strip. Leave the test strip 
in place until the result is displayed on the screen.

Bad sample:

 Insert a new strip and rerun the test.  

If the error code persists, perform the test using  
an alternate test strip vial or alternate method.

Replace strip:

 Occurs after insertion of strip or  

occurs during analysis. Insert another strip and  
retest. If the error code persists, perform the 
test using an alternate test strip vial or alternate 
method.

Temperature:

 Meter will only work within the  

temperature range of 1°C–40°C. Return the meter 
to an environment within the specified  
temperature range 1°C–40°C.

8127-02EN 04/2021.

Transfer failed:

 Server refuses to allow dialog with 

meter, or Connection to server was broken. Please 
check the network settings, status of your network, 
or contact your administrator for assistance.