System Setup Regular
Alternating Mattress Replacement System
Auto Quick Setup Guide
1
2
3
4
40 mins
20 mins
Remove mattress roll
from carry bag and lay on
bed base at foot end.
Try to avoid hanging the
control unit in an exposed
position if a lower hang
point is available.
Connect air hose to control
unit (Connector can attach
either way around)
Connect Sensor Jack to its
Port underneath air hoses.
(Control Unit will NOT
operate without sensor!)
Connect male power cord to control unit as shown.
Connect breakaway power
cord into wall socket.
Turn on control unit at
switch near power socket.
Mattress will take up to
40 minutes to inflate.
Press and hold Unlock
button for 2 seconds to
activate control panel.
Engage Care Mode for
patient transfer onto
mattress.
Care Mode will auto
disengage after 20 mins.
2 secs
Control Unit Operation
5
The Theraflow Automatic Sensor System is designed to
automatically adjust to patient weight, position & profile
Press the pressure setting button at left
All three lights will flash. The system takes
8 minutes to adjust to the patient’s weight
Locked Mode:
Press for 2 seconds to unlock all controls
Active Mode:
Default Mode for active alternation therapy
Care Mode:
Maximum pressure for patient transfers
Will default to Active Mode after 20 minutes
Static Mode:
Engage for transport or meals for comfort
Will change back to alternation after 1 hour
CPR
Ensure CPR tag is closed!
Care & Cleaning
Coverlet Cleaning
The Hyperstretch Coverlet can be machine washed and dried by following
the printed instructions on the top side of the coverlet itself - reprinted
below.
Coverlet Removal and Replacement
The coverlet can be removed for seperate cleaning by simply removing the
top cover and then unziping away from the clear PVC flaps on each side.
With the Bariatric model you need to be unzip the coverlet skirt from head
end to 360 degrees.
12
9
On-site Cleaning
The Theraflow alternating mattress replacement system features fully
welded outer surfaces. When necessary the top cover and base can be
cleaned and disinfected on site once the patient has been removed from the
mattress.
The following on-site cleaning procedure is recommended for top cover and
control unit. Note that a summary of the below is printed at the foot end
flap of the top cover. Do not immerse the control unit in water!
1. Ensure gloves are worn and all disinfection and occupational health and
safety protocols of the facility are adhered to.
2. Wipe down with a clean cloth using a disinfectant solution comprising of
hand hot water and a neutral detergent or with a sodium hypochlorite
solution (0.1% or 1000 parts per million available chlorine.
Proprietary disinfectants may be used provided manufacturer’s
instructions are followed.
3. All cleaning agents and disinfectants must be thoroughly rinsed off and
the surface dried before storage or re-use.
Failure to do this may result in the accumulation of reagent that could
damage the polyurethane coating, react with the bed frame or negate
the bio-compatibility results of the fabric.
Machine Washing
The system is designed as a “mattress within a mattress” so that the base
and top cover can be easily separated from the cell section which is fitted
within its own inner base.
Machine wash the outer mattress top cover as follows;
Wash at a temperature up to 71°
(160°F), using normal detergents
In washing machines, it may be difficult to wet out full covers.
Correspondingly, spinning and tumbling may not remove water trapped
between layers. It may be helpful to interrupt the washing or drying cycles
to alleviate this.
Machine Drying:
Drying may be achieved by hanging out, spinning or tumbling at
temperatures up to 130°C
(266°F) - Do not mangle.
Care & Cleaning
Control Unit Functions
Unlocking the system
After start up, or after 5 minutes without user input, the system will
automatically lock to prevent tampering. The Unlock LED will light up
to indicate this. To change any settings, simply press AND HOLD the
Unlock Button for 2 seconds to activate controls.
Positioning or transferring patient on bed
To position the patient on the mattress or to generally manoeuvre
or transfer the patient press the Care Button. This will inflate the
mattress to maximum pressure and provide a stable surface. The
system will automatically revert to the previous mode after 20 minutes.
Automatic Patient profiling
If the mattress head section is raised more than 25
degrees, a sensor will automatically engage Profile
Mode and extra air will be added to the system for
greater patient support and to prevent bottoming out.
Comfort Auto Control
The system is designed to automatically adjust
to patient weight, position and profile.
Once the patient is placed on the mattress
press the comfort button. All three lights will
flash and the system will take 8 minutes to adjust to the patient.
PRESSURE
POWER
TRANSPORT
LOCK/UNLOCK
THERAPY
PATIENT
TRANSFER