System Setup Regular

Alternating Mattress Replacement System

Auto Quick Setup Guide

1

2

3

4

40 mins

20 mins

Remove mattress roll 

from carry bag and lay on 

bed base at foot end.
Try to avoid hanging the 

control unit in an exposed 

position if a lower hang 

point is available.

Connect air hose to control 

unit (Connector can attach 

either way around)

Connect Sensor Jack to its 

Port underneath air hoses.

(Control Unit will NOT 

operate without sensor!)

Connect male power cord to control unit as shown.

Connect breakaway power 

cord into wall socket. 
Turn on control unit at 

switch near power socket.
Mattress will take up to

40 minutes to inflate.

Press and hold Unlock 

button for 2 seconds to 

activate control panel.

Engage Care Mode for 

patient transfer onto 

mattress. 
Care Mode will auto 

disengage after 20 mins.

2 secs

Control Unit Operation

5

The Theraflow Automatic Sensor System is designed to 

automatically adjust to patient weight, position & profile  

 

 

Press the pressure setting button at left 

 

 

All three lights will flash. The system takes

  

 

8 minutes to adjust to the patient’s weight

Locked Mode:

 Press for 2 seconds to unlock all controls

Active Mode:

  Default Mode for active alternation therapy

Care Mode:

  Maximum pressure for patient transfers 

 

 

Will default to Active Mode after 20 minutes 

Static Mode:

  Engage for transport or meals for comfort

  

 

Will change back to alternation after 1 hour

CPR

Ensure CPR tag is closed!

Care & Cleaning

Coverlet Cleaning

The Hyperstretch Coverlet can be machine washed and dried by following 

the printed instructions on the top side of the coverlet itself - reprinted 

below. 

Coverlet Removal and Replacement

The coverlet can be removed for seperate cleaning by simply removing the 

top cover and then unziping away from the clear PVC flaps on each side. 

With the Bariatric model you need to be unzip the coverlet skirt from head 

end to 360 degrees.

12

9

On-site Cleaning

The Theraflow alternating mattress replacement system features fully 

welded outer surfaces.  When necessary the top cover and base can be 

cleaned and disinfected on site once the patient has been removed from the 

mattress.

The following on-site cleaning procedure is recommended for top cover and 

control unit.  Note that a summary of the below is printed at the foot end 

flap of the top cover.  Do not immerse the control unit in water!

1. Ensure gloves are worn and all disinfection and occupational health and 

safety protocols of the facility are adhered to.

2. Wipe down with a clean cloth using a disinfectant solution comprising of 

hand hot water and a neutral detergent or with a sodium hypochlorite 

solution (0.1% or 1000 parts per million available chlorine. 

Proprietary disinfectants may be used provided manufacturer’s 

instructions are followed.

3. All cleaning agents and disinfectants must be thoroughly rinsed off and 

the surface dried before storage or re-use.   

Failure to do this may result in the accumulation of reagent that could 

damage the polyurethane coating, react with the bed frame or negate 

the bio-compatibility results of the fabric.

Machine Washing

The system is designed as a “mattress within a mattress” so that the base 

and top cover can be easily separated from the cell section which is fitted 

within its own inner base.

Machine wash the outer mattress top cover as follows;

Wash at a temperature up to 71°

 

(160°F), using normal detergents

In washing machines, it may be difficult to wet out full covers.  

Correspondingly, spinning and tumbling may not remove water trapped 

between layers.  It may be helpful to interrupt the washing or drying cycles 

to alleviate this.

Machine Drying:

Drying may be achieved by hanging out, spinning or tumbling at 

temperatures up to 130°C

 

(266°F) - Do not mangle.

Care & Cleaning

Control Unit Functions

Unlocking the system 

After start up, or after 5 minutes without user input, the system will 

automatically lock to prevent tampering.  The Unlock LED will light up 

to indicate this.  To change any settings, simply press AND HOLD the 

Unlock Button for 2 seconds to activate controls.

Positioning or transferring patient on bed 

To position the patient on the mattress or to generally manoeuvre 

or transfer the patient press the Care Button.   This will inflate the 

mattress to maximum pressure and provide a stable surface.   The 

system will automatically revert to the previous mode after 20 minutes.

Automatic Patient profiling 

If the mattress head section is raised more than 25 

degrees, a sensor will automatically engage Profile 

Mode and extra air will be added to the system for 

greater patient support and to prevent bottoming out. 

Comfort Auto Control

   The system is designed to automatically adjust  

   to patient weight, position and profile.   

   Once the patient is placed on the mattress  

   press the comfort button. All three lights will 

flash and the system will take 8 minutes to adjust to the patient. 

PRESSURE

POWER

 TRANSPORT

LOCK/UNLOCK

THERAPY

PATIENT

TRANSFER

Содержание Theraflow MRS-TFW-001 Standard

Страница 1: ...Operating Manual Theraflow Alternating Mattress Replacement Systems Model MRS TFW 001 Standard...

Страница 2: ...that cannot be excluded under the Australian Consumer Law You are entitled to a replacement or refund for a major failure and compensation for any other reasonably foreseeable loss or damage You are a...

Страница 3: ...flick open with small flat head screwdriver and replace fuse if necessary Only use fuse type 10Amp 250Volt 1 Ensure that the CPR hose connector is properly attached to the side of the control unit 2...

Страница 4: ...steps to remove the system from the bed 1 Ensure gloves are worn and all disinfection and occupational health and safety protocols of the facility are adhered to 2 If necessary clean and disinfect fol...

Страница 5: ...are worn and all disinfection and occupational health and safety protocols of the facility are adhered to 2 Wipe down with a clean cloth using a disinfectant solution comprising of hand hot water and...

Страница 6: ...of this manual PRESSURE POWER TRANSPORT LOCK UNLOCK THERAPY PATIENT TRANSFER Control Unit Functions 10 11 Operation Mode Selection The control unit has three modes of operation Active Dynamic Mode is...

Страница 7: ...re and provide a stable surface The system will automatically revert to the previous mode after 20 minutes Automatic Patient profiling If the mattress head section is raised more than 25 degrees a sen...

Страница 8: ...t as there is risk of electric shock Control units should only be opened by approved technicians or warranty will be voided Avoid blocking the air intake filter at the rear of the control unit Ensure...

Страница 9: ...gned to automatically adjust to patient weight position profile Press the pressure setting button at left All three lights will flash The system takes 8 minutes to adjust to the patient s weight Locke...

Страница 10: ...ith instructions provided with the Product 4 The Product consisting of only Patient Support Systems Pty Ltd approved parts Limited Warranty System Summary System Description 4 System Components 5 Reco...

Страница 11: ...any indirect special incidental or consequential damages even if Patient Support Systems Pty Ltd or its authorized dealers have been advised of the possibility of such damages 2 The sole remedy for br...

Отзывы: