NOISE:
Designed to operate in high noise environments (400 V line spikes, and RF immunity to high power
UHF portable transmitters In close proximity).
DISPLAYS:
4 1/2 digit with minus sign & decimals (positive values indicated by no minus sign).
Bright .56” Red LED
Fixed or floating decimal (3 places)
Number range
±
0.000 to
±
19999
Three (3) LED function indicator lamps
OUTPUTS:
Analog
0-20mA or 4-20mA field selectable, output consists of 1000 .02mA steps.
750 ohm load maximum.
Span pot adjustable
±
10%. Zero and full scale set into memory in engineering units.
0-10 VDC or 0-5 VDC obtained by installing a 500 ohm or 250 ohm precision, low temperature
coefficient resistor across the driven load.
Relay Setpoints:
2 Relays standard
SPDT (form C) contact rating (resistive) 250 VAC 6 amp (1800 VA), 28 VDC 6 amp (170W).
Selectable Relay Logic: Energize or de-energize above or below setpoint, auto-reset with
hysteresis selectable 0-99% of SP In 1% steps, latching (reset by pushbutton behind front panel
door).
Accuracy:
(including temp. variations)
Digital
±
.03% typical (
±
.05% max.) and
±
1 least significant digit
Analog
±.
3% of range
Response Times:
Display updated approx. every 1/2 sec. based on latest available input
measurement(s).
Analog & Relay Outputs
updated at a variable rate depending on the frequency. The typical
& maximum response times are:
Above 100 Hz =100 milliseconds typical
200 milliseconds maximum
2 Hz to100 Hz = 2 30 milliseconds typical
6 30 milliseconds maximum
Below 2 Hz = Measurements considered zero.
For values computed from both signal inputs, a new computed value is updated each time
either signal completes a measurement.
Range of Normalization
(linear or inverse only)
Input frequencies A & B may be normalized by a number from .5000 x 10
-7
to 2.000 x 10
+7
.
Normalization is entered in the form: + -1XXXX
+ -1XXXX
Additional display normalization range
±
.001 to 19999.
Constant Storage:
Retained In EAROM and may each be altered 1000 or more times.
Electrical References:
Circuit Common is isolated from AC power, AC ground and case. DC power, analog output and
serial output are referenced to circuit common. Passive inputs are balanced. Active pickup inputs
are referenced to circuit common.
CAUTIONARY NOTE:
Attempts to repair the TACHTROL
®
3 in the field are NOT RECOMMENDED. Should
a problem occur, call: Deca Systems, Inc.
Baton Rouge, LA
225-273-2770