5. TECHNICAL SPECIFICATIONS
70 ± 5 seconds
Braking Time
Ambient Temperature
5 - 40 °C
Permissible Relative Moisture
<80%
Size (W x D x H)
Weight
Input Power
330 X 265 X 161 mm
4.3 kg Approx (with rotor)
Power Consumption
132 W
24 V 5.6 A
6. SAFETY PRECAUTIONS
Ÿ
Never use the centrifuge in any manner not specified in this manual.
Ÿ
Using equipment in any manner not specified in this manual or by the
manufacturer, will void the warranty.
Ÿ
Never move the centrifuge while the rotor is spinning.
Ÿ
The rotor and the rotor lid must always be securely fastened. If the centrifuge
makes unusual noise during operation, the rotor or rotor lid fit needs to be checked.
Switch OFF the device immediately by pressing STOP, check rotor fit & fasten it well.
Ÿ
The rotors must be loaded symmetrically. Each tube should be counter balanced by
another tube of the same weight.
Ÿ
Do not use the centrifuge or rotor that have not been correctly installed or
damaged.
Ÿ
Repairs must only be performed by authorized service technician.
Ÿ
Using incorrect rotors & wrong spare parts will void the warranty.
Ÿ
Centrifuge may be used for the specified applications only. It must not be operated
in a hazardous or flammable environment and must not be used to centrifuge
explosive or highly reactive substances.
Ÿ
If liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned
carefully and properly before being used again.
Ÿ
Prior to centrifugation, the tubes should be visually inspected for material damage.
Damaged tubes may not be centrifuged. This is because broken tubes can, in
addition to sample loss, create imbalance which can result in further damage
to the centrifuge and accessories.
Ÿ
The capacity of 24 capillary tubes must not be exceeded as it is the maximum
capacity.
Ÿ
Do not lean on the equipment. It may damage the equipment or the harm the
operator.
Ÿ
When moving the centrifuge from a cold room to a normal room, run the centrifuge
for 30 minutes beforehand in the cold room to avoid condensation. Alternately,
allow it to warm up in the lab for at least 3 hours before use, but do not plug in the
centrifuge in order to prevent possible damage by condensation.