The
output format
can also be set in the function output mode ´
OM
´. The three
options can be selected by using the key CH
▲
. For identification purposes the
abbreviations ´
n
´ or ´
t
´ follow the output channel ´
U
´:
Abbr.
Output Format
U
measured values in list format
Un
measured values in column format, side by side
Ut
measured values in spreadsheet (table) format
8.5.3 Manual Data Output
In addition to the list of measured values data such as max/min values, the
sensor and instrument programming can, similar to an output via built-in
printer, be provided as output via interface by using the key
START/STOP
(s.8.3).
All other function values that have been selected by using the keys
MEASURE
or
FUNCTION
can be either printed out by using interface commands, or by using
an additional trigger cable (accessory ZA 1000-ET, variant 3, see man. 6.10.9).
The trigger cable must be connected to socket A1 and, with the external key,
the outputs providing the following printouts can be triggered.
The following functions can also be output by using the key
START/STOP
:
Function
Ab Ky Printout
OUTPUT MODE
BAUD RATE
(sensor
programming)
OM
BR
F
F
AMR ALMEMO 6290-7
see man. 6.2.3
CH RANGE LIM-MAX LIM-MIN BASE D FACTOR EXP AVG COMMENT
01:NiCr +0123.4 - - - - - - °C 1.0350 E+0 - - - Designation
02:NiCr - - - +0012.0 - - - °C - - - E+0 CONT Water
MEAS. CYCLE: 00:00:30 S S0501.3 F0204.7 AR W010 C--U-
PRINT CYCLE: 00:10:00 Un 9600 bd
ATM. PRESSURE
DEVICE ADDRESS
(instrument
programming)
mb
A
F
DEVICE: G00 M11 A01 P03/11/00
see man. 6.2.5
A.PRESSURE: +01013. mb
CJ-TEMP: +0023.5 °C
U-SENSOR: ! 12.5 V
HYSTERESIS: 10
CONFIG: FCRD-S-- -L--
ALARM: -1-3
A1: DK0 Un
A2: AK1
MAX/MIN VALUE
(all meas. values)
MH
ML
CH
CH MEAS.VAL MAXVAL MINVAL AVG COUNT
01:+0023.0 +0025.0 +0019.0 - - - - - -
s. man. 6.4.4
MEMORY
MY
CH
MEMORY: - - - -
see man. 6.9.2
The following functions can only be output with the trigger cable:
MEAS. VALUE
Dim
CH
12:34:00 01: +0023.5 °C
PRINT CYCLE
PC
F
PRINT CYCLE: 00:06:00
PAPER FEED
PF
F
FEED: 130.00mm/h DIVISION: 5 min
TIME
TM
F
TIME: 12:34:00
DATE
DA
F
DATE: 01.02.99
RANGE
R
F
01:NiCr +0123.4 -0012.0 +0000.0 °C 1.0000 E+0 - - -
ALMEMO
6290-7B
40
Digital Data Output