12. Device configuration
29
different ways or even switched off in cases where there is a risk of accidentally
activating relative measuring by zero-setting the measured value.
Loc
parameter :
0 The offset is saved in RAM, base or zero-point - depending on locking
1 The offset is saved in RAM only.
2 Relative measuring is locked
12.5
Atmospheric pressure compensation
Measured variables dependent on the ambient atmospheric pressure (see Man-
ual 6.3.3 Measuring range list ´with PC´) may, in the event of large deviations
from normal pressure (1013 mbar), involve certain measuring errors.
e.g. Error per 100 mbar
Compensation range
Rel.humidity psychrometer
approx. 2% 500 to 1500 mbar
Mixture ratio, capacitive
approx. 10 % Vapor pressure VP up to 8 bar
Dynamic pressure
approx. 5% 800 to 1250 mbar (error < 2%)
O
2
saturation
approx. 10% 500 to 1500 mbar
It is advisable therefore, especially when taking measurements at appreciable
heights above sea level to take due account of the atmospheric pressure (ap-
prox. -11 mbar / 100 meters above mean sea level, MSL).
The appropriate atmospheric pressure can be entered in parameter
'mb'
either
in device programming or it can be measured using an atmospheric pressure
sensor (reference sensor with designation ´*P´, see Manual 6.7.2).