10. Measuring operations
21
Transducers
Sensor / con-
nector
Measuring range
Units Abbre-
viation
Fe-CuNi (L)
ZA 9000-FSL
-200.0... +900.0
°C
FECO
Fe-CuNi (J)
ZA 9000-FSJ
-200.0...+1000.0
°C
IrCo
Cu-CuNi (U)
ZA 9000-FSU
-200.0... +600.0
°C
CUCO
Cu-CuNi (T)
ZA 9000-FST
-200.0... +400.0
°C
CoCo
PtRh10-Pt (S)
FS Axxx
0.0...+1760.0
°C
Pt10
PtRh13-Pt (R)
ZA 9000-FSR
0.0...+1760.0
°C
Pt13
PtRh30-PtRh6 (B)
ZA 9000-FSB
+400.0...+1800.0
°C
EL18
Au-FeCr
ZA 9000-FSA
-270.0... +60.0
°C
AUFE
Ntc Typ N
FN Axxx
-30.00...+125.00
°C
Ntc
Millivolt
ZA 9000-FS0
-10.000...+55.000
mV
U 55
Millivolt 1
ZA 9000-FS1
-26.000...+26.000
mV
U 26
Millivolt 2
ZA 9000-FS2
-260.00...+260.00
mV
U260
Volt
ZA 9000-FS3
-2.0000...+2.6000
V
U2
.
60
Difference millivolt
ZA 9000-FS0D
-10.000...+55.000
mV
d 55
Difference millivolt 1
ZA 9000-FS1D
-26.000...+26.000
mV
d 26
Difference millivolt 2
ZA 9000-FS2D
-260.00...+260.00
mV
d260
Difference volt
ZA 9000-FS3D
-2.0000...+2.6000
V
d2
.
60
Sensor voltage
any
0.00...20.00
V
UbAt
Milliampere
ZA 9601-FS1
-26.000...+26.000
mA
I032
Percent (4-20mA)
ZA 9601-FS2
0.00... 100.00
%
P420
Ohm
ZA 9003-FS
0.0... 500.0
Ohn
Frequence
ZA 9909-AK1
0... 32000
Hz
FrEq
Pulse
ZA 9909-AK2
0... 65000
PULS
Digital input
ZA 9000-EK2
0.0... 100.0
%
Inp
Digital interface
ZA 9919-AKxx
-65000... +65000
diGi
Snap-on head normal 20
FV A915-S120
0.30... 20.00
m/s
S120
Snap-on head normal 40
FV A915-S140
0.40... 40.00
m/s
S140
Snap-on head micro 20
FV A915-S220
0.50... 20.00
m/s
S220
Snap-on head micro 40
FV A915-S240
0.60... 40.00
m/s
S240
Macro
FV A915-MA1
0.10... 20.00
m/s
L420
Water-Micro
FV A915-WM1
0.00... 5.00
m/s
L605
Dynamic press., 40 m/s with TC and PC
FD A612-M1
0.50... 40.00
m/s
L840
Dynamic press., 90 m/s with TC and PC
FD A612-M6
1.00... 90.00
m/s
L890
Relative air humidity, capacitive
FH A646
0.0... 100.0
%H
°orH
Relative air humidity, cap., TC
FH A646-C
0.0... 100.0
%H
HcrH
Relative air humidity, cap., TC
FH A646-R
0.0... 100.0
%H
H rH
Mixture ratio, capacitive with PC
FH A646
0.0 ... 500.0
g/k
H AH
Dew-point temperature, cap.
FH A646
-25.0... 100.0
°C
H dt
Partial vapor pressure, cap.
FH A646
0.0 ...1050.0
mb
H UP
Enthalpy, capacitive with PC
FH A646
0.0 ... 400.0
kJ
H En
Humid temperature
FN A846
-30.00 ... +125.00
°C
P Ht
Relative humidity, psychr. with PC
FN A846
0.0 ... 100.0
%H
P RH
Mixture ratio, psychrometric with PC
FN A846
0.0 ... 500.0
g/k
P AH
Dew-point temp., psychrometric with PC
FN A846
-25.0 ... +100.0
°C
P dt
Partial vapor pressure, psychr. with PC
FN A846
0.0 ...1050.0
mb
P UP
Содержание ALMEMO 2490-1A
Страница 32: ...32 14 DECLARATION OF CONFORMITY...