14. Sensor programming
select the required input channel, and then enter the range. (see 11.1) When
the input for the new measuring range is confirmed all programming values for
that input channel will be deleted.
Function range selection
RANGE:
D dv
To accept the selection of all possible measuring channels press
<MALL>
To deactivate a channel press
<CLR>
To reactivate a channel press
PROG
,
PROG
Programming the range is as for data input s.9.5
PROG
,
... ,
PROG
In the input window all possible abbreviations appear one after the other, as
listed in the following table :
RANGE:
D dv
And a help window for identifying the sensors
14.4 Functions of the Pt100 psychrometer FPA8363P3
14.4.1 Measuring ranges
The following table lists the measuring ranges available on Pt100 psychrome-
ter FPA8363P3. Factory default settings are marked with an asterisk ´*´. Hu-
midity variables must be programmed in sensor M1 to channels M11, M21,
M31 and atmospheric pressure, differential, or supply voltage to channels M02
to M32. (see 8.3) Commands for entering this via an interface using a termi-
nal are also listed.
Measuring function
Sensor
Meas. Range Unit
Display
Com-
mand
Pt100-3
ITS90
0.001°C
FPA903L0250
-200.000...+400.000 °C
P314
B96
*Pt100-3
ITS90
HT
FPA8363P3
M0
-200.000...+400.000 °C
P314
B96
*Pt100-3
ITS90
TT
FPA8363P3
M1
-200.000...+400.000 °C
P314
B96
*Atmospheric pressure AP int. Device channel M2
300.0... 1100.0 mbar
AP
B86
*Rel. Humidity 0.01%H
FPA8363P3
M1
0.00... 100.00 %H
P2RH
B-1O
Dew-point temp. 0.01°C
FPA8363P3
M1
-64.80... 100.00 °C
P2DT
B-11
Mixture MH, r
FPA8363P3
M1
0.0... 6500.0 g/kg
D MH
B-O4
Abs. Humidity AH, dv
FPA8363P3
M1
0.0... 596.3 g/m
3
D AH
B-O5
Vapor pressure VP, e
FPA8363P3
M1
300... 1100.0 mbar
D VP
B-O6
Enthalpy En, h
FPA8363P3
M1
0.0... 6500.0 kJ/kg
D En
B-O7
Differential
Device channel M2
Diff
B71
Supply voltage
Device channel M2
0.0... 26.0 V
Batt
B14
14.4.2 Multi-point adjustment
The sensor’s measuring accuracy can be optimized by storing in the connector
EEPROM a correction curve for the Pt100 dry temperature and humid temper-
ature sensors. In a calibration laboratory any deviations from the ideal setpoint
can be measured at various points; then, with the AMR-Control, this informa-
26 ALMEMO
®
1036
Sensor FPA836-3P3
abs. Humidity
0.0 ... 596.3 g/¦