Symptom  C:  Unit  reads  “MOISTURE 
BELOW  LIMIT”  or  “MOISTURE  ABOVE 
LIMIT.”

Solution 1:

 Grain may be too wet or dry 

to test. Check moisture limit guidelines in 
Operating Instructions.

NOTE: Moisture limits are only guide-
lines.

Symptom  D:  Unit  reads  “TEST  CELL 
ERROR”.

Solution 1:

 Fill cup may still be inside the 

test cell. Remove it and try again.

Solution  2:

  Test  cell  still  has  grain  in  it. 

Remove all grain.

SERVICE

TO REPLACE BATTERIES

The tester is supplied with two 9 volt, alka-
line batteries. The right battery (A) powers 
the  backlight  circuit.  The  left  battery  (B) 
powers the system.

At any time, select BATTERY from main menu 
to display the condition of both batteries.

The  tester  will  properly  operate,  except 
backlighting,  when  there  is  no  backlight-
ing battery and if this battery is low.

NOTE:  If  the  backlighting  battery  is 
fresh  and  the  system  battery  needs 
replacing, the backlight battery can be 
used to operate the system, by moving 
it to the system battery location.

Remove  the  battery  door  and  the  old 
battery(s).

Insert new battery(s) and replace door.

TO CLEAN MOISTURE TESTER

Remove cap and wipe out inside of mois-
ture tester cell with a dry paper towel.

ACCESSORIES

OPTIONAL CARRYING CASE

(P/N 20101)
The carrying case is constructed of pad-
ded vinyl to protect the tester. A zippered 
bottom,  clear  vinyl  window  and  Velcro

®

 

sealing  flap  permit  tester  operation  with 
tester in carrying case. Velcro is a regis-
tered trademark of VELCRO Industries.

OPTIONAL SOFTWARE

(P/N 20107)
This tester communicates to a PC through 
the RS232 serial port, like some commer-
cial  testers  do.  Software  and  grain  table 
upgrades can be downloaded through the 
RS232 interface.

See the software documentation for down-
loading.

REPLACEMENT PARTS

Tester Cap Replacement:
 

Order P/N 20103

Grain Scoop Replacement:
 

Order P/N 20104

Battery Door Replacement:
 

Order P/N 20105

Serial Port Door Replacement:
 

Order P/N 20106

DOCU-M0111  3-09

Содержание Farmex GMT2

Страница 1: ...ne year from date of retail purchase This warranty does not cover the batteries or damage resulting from mis use neglect accident or improper installation or maintenance This warranty does not apply t...

Страница 2: ...ecting the Commercial Tester Option 2 Selecting the Grain Option 3 Operation General Information 3 Operational Messages 3 Moisture Limit Guidelines 4 Operating Conditions 5 Moisture Test 6 Calibration...

Страница 3: ...tive cover and discard BATTERY CONNECTION The tester is supplied with two standard 9V alkalinebatteries Thesystembatteryprovides power for all electronics and the backlight battery powers the LCD disp...

Страница 4: ...the option required and press the ENTER button SELECTING THE BATTERY OPTION Selecting the BATTERY option will display a screen that indicates the power remain ing in the SYSTEM battery and the LIGHT...

Страница 5: ...urve that is based on the moisture read ings of a commercial grain elevator mois ture tester The accepted industry practice is for farmers to sell their grain according to the readings given by their...

Страница 6: ...4...

Страница 7: ...compaction in the test cell and allows a fixed com paction setting to be preset to establish a trip point at which a test can begin The MoistureMATCH will automatically begin testing the sample compa...

Страница 8: ...iner such as a zip lock bag or jar MOISTURE TEST The fill cup must be removed from the test cell The cap can be on or off while the tester performs self calibration Inspect the test cell and make sure...

Страница 9: ...ts are required press ENTER remove the cap and complete ly empty the test cell Press ENTER and the tester will return to the GRAIN dis play ready to begin a new test IMPORTANT A NEW SAMPLE MUST BE USE...

Страница 10: ...een the commercial tester reading and the MoistureMATCH 4 The maximum offset allowed at that moisture point This is the largest offset that a user can enter Use the UP DOWN arrow to enter the mois tur...

Страница 11: ...r short er time Do NOT hold button down Solution 2 Check batteries for 7 5 volts or higher Replace as necessary Solution 3 Remove battery access door from bottom of tester Remove batteries and ensure...

Страница 12: ...replacing the backlight battery can be used to operate the system by moving it to the system battery location Remove the battery door and the old battery s Insert new battery s and replace door TO CLE...

Отзывы: