VHS-4 Diffusion Pump

DRAFT 8/7/02

Pressurization Hazards 

Large vacuum pumps and their components are 
designed for vacuum service; they are not designed 
to be pressurized which could cause them to burst 
possibly expelling shrapnel at lethal velocities. Seri-
ous accidents have been caused by intentional pres-
surization of vacuum systems and their components. 

Never pressurize any part of a vacuum system for 
test or any other purpose.

Always provide pressure relief when designing 
diffusion pumps into systems and ensure that 
pressure relief motion is limited to safe 
envelopes. 

Never permit the hazards in Table 3 to develop. 

Pressure Relief Devices: Systems larger than 
10 inches, such as this pump, must be designed with 
pressure relief devices to provide safe pressure relief 
from internal explosions. Always recognize that 
safety devices can fail or malfunction; provide redun-
dant protection by installing devices having different 
failure modes, failure mechanisms, and failure 
causes. Be certain that exhaust duct materials are 
capable of withstanding the corrosivity, temperature, 
and pressure of exhausted products.

Dangerous Substances

Chemical Dangers of Acetone and Alcohol: Diffusion 
pumps are typically cleaned with acetone or alcohol. 
Acetone, alcohol, and most other solvents are irri-
tants, narcotics, and depressants, and/or carcino-
genic. Their inhalation and ingestion may produce 
serious effects. Even absorption through the skin can 
result in moderate toxicity. Always ensure that clean-
ing operations are performed in large, well-ventilated 
rooms. Use of self-contained breathing apparatus 
may be necessary depending upon the solvent type 
and vapor concentration in surrounding air.

Poisonous and Corrosive Compounds: When pump-
ing poisonous, reactive, and/or corrosive gas, vapors, 
or chemicals, proper operation and regeneration do 
not always ensure that all hazardous materials have 
been totally removed. If hazardous gas, vapors, 
chemicals, or combustible mixtures are pumped, suf-
ficient quantities may exist during operation or 
remain after regeneration to cause severe injury or 
death. 

Pump Fluids: Overheating the pump fluid, exposing it 
to air or reactive materials, or overpressurizing it 
above the normal operating range, approximately 
1x10

-3

 Torr (1.3x10

-3

 mbar) decomposes the fluid 

and possibly makes it toxic. This is especially true of 
backstreamed mechanical pump oils which are more 
volatile (unstable). Overheating of accidentally intro-
duced or backstreamed mechanical pump oils can-
not be protected against by thermal switches which 
are set for diffusion pump oil.

Process Gasses: These gasses are frequently toxic, 
flammable, corrosive, explosive, or otherwise reac-
tive. Varian has no control over the types of gasses 
passing through the user’s diffusion pump as these 
are entirely under the control of the process user 
and/or the hardware systems integrator. Since these 
gasses can cause serious injury or death, it is very 
important to plumb the exhaust of the pump to the 
facility’s hazardous gas exhaust system which incor-
porates appropriate filters, scrubbers and similar 
components to ensure that the exhaust meets all air 
and water pollution control regulations.

Table 3

Pressurization Hazards

Prohibited Action

Result

Do not block inlet or 
vent of liquid nitrogen 
trap and lines

LN

2

 trap and/or lines 

burst

Do not close isolation 
valves at inlet and 
discharge of main water 
cooling coils when 
pump is reheated

Water turns to steam and 
bursts coils

Do not pressurize the 
pump body

Body of pump bursts

Do not make a hole 
through the vacuum wall

Loss of structural 
integrity of wall

Содержание VHS-4

Страница 1: ...m technologies VHS 4 Diffusion Pump Part No 699901021 Rev No D Date October 2000 NOTICE This document contains references to Varian Please note that Varian Inc is now part of Agilent Technologies For more information go to www agilent com chem ...

Страница 2: ...D R A F T 8 7 0 2 Copyright 2000 Varian Vacuum Technologies VHS 4 Diffusion Pump ...

Страница 3: ...ng standard s or other normative documents auf das sich diese Erklärung bezieht mit der den flogenden Norm en oder Richtlinie n übereinstimmt auquel se réfère cette déclaration est conforme à la auz norme s ou au x document s normatif s al que se refiere esta declaración es conforme a la s norma s u otro s documento s normativo s waamaar deze verklaring verwijst aan de volende norm en of richtlijn...

Страница 4: ...serious risk It is the responsibility of the user to maintain safe operating conditions at all times Varian assumes no liability for personal injury or damage resulting from operation or service of the equipment Varian has no control over the use of this equipment and is not responsible for personal injury or damage resulting from its use The safe use and disposal of hazardous or potentially hazar...

Страница 5: ... improper use by untrained persons non authorized interference with the equipment or any action contrary to that proved for by specific national standards This manual uses the following standard safety protocol WARNING The warnings are for attracting the attention of the operator to a particular procedure or practice which if not followed correctly could lead to serious injury CAUTION The cautions...

Страница 6: ...VHS 4 Diffusion Pump This page intentionally left blank D R A F T 8 7 0 2 ...

Страница 7: ... 4 Unpacking 4 Pump Oil Installation 4 Vacuum System Connections 4 Cooling Water Connections 5 Electrical Connections 5 Thermal Switch 6 Section II Operation 7 Startup Procedure 7 Shutdown Procedure 8 Section III Maintenance and Service 9 General 9 Periodic Inspection 9 Cleaning 9 Cold Cap Removal Installation 10 Jet Assembly Removal Installation 11 Heater Replacement 12 Adding Pump Fluid 12 Secti...

Страница 8: ...VHS 4 Diffusion Pump This page intentionally left blank D R A F T 8 7 0 2 ...

Страница 9: ...vide an interlock to ensure that the power supply to the pump heater cannot be activated if the foreline pump is not running and or the pressure in foreline is above 0 5 Torr 0 38 mbar Overtemperature Fit temperature sensors and pump fluid level sensors with feedback to an interlock on the heater power supply Insufficient water flow through the main cooling coils Use water flow sensor and feedback...

Страница 10: ...s with clean dry oil free compressed air Oxygen and other strong oxidizers are even more dangerous than air Certain conditions of temperature and pressure can cause a combustible mixture to explode The larger the diffusion pump the greater the risk of explosion and the greater the risk of dam age and injury Never operate large diffusion pumps utilizing hydrocarbon oils without a complete safety an...

Страница 11: ...nous and Corrosive Compounds When pump ing poisonous reactive and or corrosive gas vapors or chemicals proper operation and regeneration do not always ensure that all hazardous materials have been totally removed If hazardous gas vapors chemicals or combustible mixtures are pumped suf ficient quantities may exist during operation or remain after regeneration to cause severe injury or death Pump Fl...

Страница 12: ...oils from previous use turns to steam when the pump is reheated This steam must be allowed to escape with out contacting personnel Whenever possible design the water system with interlock valves so that power cannot be applied to the pump unless water is flow ing in the main cooling coils not Quick Cool coils High Voltages Diffusion pump heaters operate at voltages high enough to kill Design syste...

Страница 13: ...1 VHS 4 Diffusion Pump D R A F T ...

Страница 14: ...VHS 4 Diffusion Pump 2 D R A F T 8 7 0 2 ...

Страница 15: ...3 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 16: ...VHS 4 Diffusion Pump 4 D R A F T 8 7 0 2 ...

Страница 17: ...5 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 18: ...VHS 4 Diffusion Pump 6 D R A F T 8 7 0 2 ...

Страница 19: ...VHS 4 Diffusion Pump 7 D R A F T 8 7 0 2 ...

Страница 20: ...VHS 4 Diffusion Pump 8 D R A F T 8 7 0 2 ...

Страница 21: ...9 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 22: ...VHS 4 Diffusion Pump 10 D R A F T 8 7 0 2 ...

Страница 23: ...11 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 24: ...VHS 4 Diffusion Pump 12 D R A F T 8 7 0 2 ...

Страница 25: ...13 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 26: ...VHS 4 Diffusion Pump This page intentionally left blank D R A F T 8 7 0 2 ...

Страница 27: ...15 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 28: ...VHS 4 Diffusion Pump 16 D R A F T 8 7 0 2 ...

Страница 29: ...17 VHS 4 Diffusion Pump D R A F T 8 7 0 2 ...

Страница 30: ...VHS 4 Diffusion Pump This page intentionally left blank D R A F T 8 7 0 2 ...

Страница 31: ......

Страница 32: ...ian India PVT LTD 101 108 1st Floor 1010 Competent House 7 Nangal Raya Business Centre New Delhi 110 046 India Tel 91 11 5548444 Fax 91 11 5548445 Italy Varian Vacuum Technologies Via F lli Varian 54 10040 Leini Torino Italy Tel 39 011 997 9111 Fax 39 011 997 9350 Japan Varian Vacuum Technologies Sumitomo Shibaura Building 8th Floor 4 16 36 Shibaura Minato ku Tokyo 108 Japan Tel 81 3 5232 1253 Fax...

Отзывы: