Agilent Technologies Varian 3800 GC Скачать руководство пользователя страница 6

Deuterium Exchange

6

parameters may be easily controlled. In addition to auto-ionization, overfilling the trap decreases
mass resolution. Ion traps have an automatic mechanism for controlling the number of ions in the
trap, called automatic gain control, or AGC. The system does a pre-scan to determine the signal
strength and then adjusts the hold time in the trap to avoid filling the trap with too many ions.
The pre-scan signal is compared to a target ion count. The hold time is adjusted to match this
target. If there is insufficient sample to meet the target, the hold time is increased. However, for
very low sample concentrations the hold time may be impractical so a maximum hold time is
specified to avoid slowing down the data rate unduly. To avoid auto-ionization, the target ion
count and the maximum hold time may be decreased. For example, the default target ion count is
25000 and the default maximum hold time is 25000 us. To avoid auto-ionization the target ion
count can be decreased to 500 and the maximum hold time in the trap may be decreased to
2000 us or less. The method “” is set up in this way. Sample amounts can also be decreased to
cut down on the concentration of ions in the trap.

Another advantage of using decreased target ion counts and maximum hold times is resolution

is improved. The disadvantage of using decreased target ion counts and maximum hold times is
less sensitivity. Since ion trap mass spectrometers are about ten times as sensitive as quadrupole
mass analyzers, even when operating with decreased source controls, you can still have
comparable sensitivity to a quadrupole mass spectrometer.

How do you know if auto-ionization is a problem for your sample? Compare the mass spectra

on the leading and trailing edge of your chromatogram peak with the spectra near the apex. If the
M+1/M ratio changes drastically, with the spectra near the apex having much larger M+1/M
ratios, then auto-ionization is a problem. For approximate M+1/M ratios, you can then just use
the spectra from the leading or trailing edge.

Another way to tell if auto-ionization is a problem is to compare peaks in the spectrum. If your

compound has an even molecular weight, most of the fragment peaks will have odd masses.
Conversely, if your molecular weight is odd, most of the fragment ions will have even masses. If
you have good signal to noise, you should do another run with much smaller amounts of material.
However, to be careful, you should repeat your run with much smaller target ion counts and
maximum hold times, in addition to smaller amounts using the m50t200m12mw method.

Another indicator that auto-ionization is a problem is the appearance of a large number of

trailing peaks:

If your spectrum takes on this appearance, you need to dilute your sample by a factor of 50-100.

m/z

Содержание Varian 3800 GC

Страница 1: ...middle of the top icon bar just to the right of the method pull down list Find the folder for your course or research group you may need to go up a level In this folder the methods m50t200m12 m50t200m12CI or m50t200m12mw are good general methods See the notes at the end of these instructions for a discussion of molecular weight determination This method runs a temperature gradient from 50 to 200 C...

Страница 2: ... the first use flush out a syringe with methylene chloride Be sure to eject the cleaning solvent into a waste container Do not eject back into the source bottle as it will contaminate the entire bottle Flush out the syringe at least 50 times with air to remove traces of solvent 4 Head Space Analysis If you are doing head space analysis flush out the syringe with your sample vapor at least three ti...

Страница 3: ...the Cursor Display icon This icon looks like a small mass spectrum with one red peak Clicking again hides this information box Data Analysis 10 When the chromatographic peak returns to baseline or the run is completed you can display the data in the Review Process MS Data window To open your data in the Review Process MS window click on the small black icon that has a representation of the chromat...

Страница 4: ...MultiChrom window and the SatView window 16 You should now see the 2000 40 MS acquisition window If necessary click on the Show Keypad button to see the acquisition controls Check the status indicators If they are both green the system is ready for the next sample 17 Rinse the injection syringe 25 times with methanol SHUT DOWN PROCEDURE 1 Minimize the 2000 44 acquisition control window 2 Maximize ...

Страница 5: ...n M C2H2N M C2H2N The adduct peaks are usually much lower in intensity than your parent MH peak For some types of samples the normal M ion is also produced by a charge transfer process M C2H3N M C2H3N For example acetonitrile CI works well with compounds containing nitrogen atoms producing a characteristic MH ion and a confirmatory M 13 adduct ion probably from the reaction with CH3CNCH2 Saturated...

Страница 6: ...s improved The disadvantage of using decreased target ion counts and maximum hold times is less sensitivity Since ion trap mass spectrometers are about ten times as sensitive as quadrupole mass analyzers even when operating with decreased source controls you can still have comparable sensitivity to a quadrupole mass spectrometer How do you know if auto ionization is a problem for your sample Compa...

Отзывы: