Deuterium Exchange

2

9. The tuning procedure calibrates the mass axis. Choose the “FC-43 mass calibration” tuning
mode in the pull-down list box in the middle of the window. Make sure the check box to the left
of the pull-down list box is checked, and the other three check boxes are cleared. Click on Start
Auto Tune. The system will open the calibration valve on the MS. This valve allows the vapor
from a fluorinated organic liquid, FC-43, to enter the MS. The text box across the bottom the
screen shows the progress of the Auto Tune procedure. After the procedure is complete, check to
see that “Multipoint mass cal: completed” has been printed in the text box.

10. Click on the Acquisition button, at the far right of the MS control window.

RUNNING A SAMPLE

1. The status indicators should both be green and be labeled Ready and No Faults.

2. The system is ready to begin. The start of data acquisition is signaled by the mechanical
switch on the top of the injection port. When you press down on this switch with the syringe,
data acquisition will begin.

3. Remember to use only a blunt tipped, conical needle syringe, to avoid destroying the Merlin
GC inlet. Before the first use, flush out a syringe with methylene chloride. Be sure to eject the
cleaning solvent into a waste container. Do not eject back into the source bottle as it will
contaminate the entire bottle. Flush out the syringe at least 50 times with air to remove traces of
solvent.

4. Head Space Analysis: If you are doing head-space analysis flush out the syringe with your
sample vapor at least three times. Inject 2-3 uL of your sample by gently pushing the syringe
needle into the injection port on the top of the GC. Be sure to push straight down and guide the
needle in as it is very fragile and easily bent. Remove the syringe immediately after injection.

5. Solution Analysis: Inject 0.2-0.3 uL of your sample by gently pushing the syringe needle into
the injection port on the top of the GC. That’s correct—the syringe has just a tiny amount of
solution in it. Be sure to push straight down and guide the needle in as it is very fragile and
easily bent. Remove the syringe immediately after injection.

6. Note the data file name. This name will be listed in the upper-right corner of the acquisition
dialog window. It will be in a form such as “2000.40012.SMS.”

Ion Trap Mass specs are very sensitive. You must
be very careful not to overload the instrument
.
If you are injecting just gases, no liquids, use step 4.
If you are injecting liquids skip to step 5.

Содержание Varian 3800 GC

Страница 1: ...middle of the top icon bar just to the right of the method pull down list Find the folder for your course or research group you may need to go up a level In this folder the methods m50t200m12 m50t200m12CI or m50t200m12mw are good general methods See the notes at the end of these instructions for a discussion of molecular weight determination This method runs a temperature gradient from 50 to 200 C...

Страница 2: ... the first use flush out a syringe with methylene chloride Be sure to eject the cleaning solvent into a waste container Do not eject back into the source bottle as it will contaminate the entire bottle Flush out the syringe at least 50 times with air to remove traces of solvent 4 Head Space Analysis If you are doing head space analysis flush out the syringe with your sample vapor at least three ti...

Страница 3: ...the Cursor Display icon This icon looks like a small mass spectrum with one red peak Clicking again hides this information box Data Analysis 10 When the chromatographic peak returns to baseline or the run is completed you can display the data in the Review Process MS Data window To open your data in the Review Process MS window click on the small black icon that has a representation of the chromat...

Страница 4: ...MultiChrom window and the SatView window 16 You should now see the 2000 40 MS acquisition window If necessary click on the Show Keypad button to see the acquisition controls Check the status indicators If they are both green the system is ready for the next sample 17 Rinse the injection syringe 25 times with methanol SHUT DOWN PROCEDURE 1 Minimize the 2000 44 acquisition control window 2 Maximize ...

Страница 5: ...n M C2H2N M C2H2N The adduct peaks are usually much lower in intensity than your parent MH peak For some types of samples the normal M ion is also produced by a charge transfer process M C2H3N M C2H3N For example acetonitrile CI works well with compounds containing nitrogen atoms producing a characteristic MH ion and a confirmatory M 13 adduct ion probably from the reaction with CH3CNCH2 Saturated...

Страница 6: ...s improved The disadvantage of using decreased target ion counts and maximum hold times is less sensitivity Since ion trap mass spectrometers are about ten times as sensitive as quadrupole mass analyzers even when operating with decreased source controls you can still have comparable sensitivity to a quadrupole mass spectrometer How do you know if auto ionization is a problem for your sample Compa...

Отзывы: