Varian 3800/2000 GC/MS Operating Instructions

General Notes:

* All mouse functions are accomplished with the left button unless specified.

* The function of each icon can be listed by placing the cursor over the icon and waiting a second
or two. The icon label will be listed on the screen.

* Don’t use the go-away box on the 3800.44 or 2000.40 control windows. Minimize them
instead.

* Do use the go-away box for the Review/Process Data- SatView window.

Procedure:

1. Check that the helium pressure gauge on the wall reads 80 psi.

3. On the computer, note the Varian Star Toolbar. This is a wide toolbar that runs down the left
side of the screen. If the Varian Star toolbar runs across the bottom of the screen, click on the
background of the toolbar and drag it to the left. Click on the System Control icon. This icon is at
the top.

4. Pull down the Windows menu and select 3800.44 to display the GC control panel.

5. To select the desired method, click on the “Activate a Method” icon. This icon is an “open file
folder” icon in the middle of the top icon bar, just to the right of the method pull down list. Find
the folder for your course or research group; you may need to go up a level. In this folder, the
methods “m50t200m12”, “m50t200m12CI”, or “m50t200m12mw” are good general methods.
See the notes at the end of these instructions for a discussion of molecular weight determination.
This method runs a temperature gradient from 50 to 200

°

C in 12 minutes. Click Open to

continue.

6. Check the GC temperature zones, which are listed in the middle of the screen. The column
oven should be 50

°

C, the front injector 225

°

C, and the middle injector 50

°

C. We will use the

front injector for this lab. The oven temperature gradient is set for 50-200

°

C at 15

°

C/min. With

an initial hold at 50

°

C for 2.0 min, the total run time is 12 min. Check for the total run time,

which is listed as the “End Time.” Note the two status indicator lights: the Equilibrating/Ready
light and the Fault/No Fault lights should both be green before beginning a run. These status
indicators are in the upper left-hand block of the control window.

7. Minimize the 3800 control window.

8. Pull down the Windows menu and select 2000.40 to display the MS control panel.

9. Click on the Auto Tune button.

Содержание Varian 3800 GC

Страница 1: ...middle of the top icon bar just to the right of the method pull down list Find the folder for your course or research group you may need to go up a level In this folder the methods m50t200m12 m50t200m12CI or m50t200m12mw are good general methods See the notes at the end of these instructions for a discussion of molecular weight determination This method runs a temperature gradient from 50 to 200 C...

Страница 2: ... the first use flush out a syringe with methylene chloride Be sure to eject the cleaning solvent into a waste container Do not eject back into the source bottle as it will contaminate the entire bottle Flush out the syringe at least 50 times with air to remove traces of solvent 4 Head Space Analysis If you are doing head space analysis flush out the syringe with your sample vapor at least three ti...

Страница 3: ...the Cursor Display icon This icon looks like a small mass spectrum with one red peak Clicking again hides this information box Data Analysis 10 When the chromatographic peak returns to baseline or the run is completed you can display the data in the Review Process MS Data window To open your data in the Review Process MS window click on the small black icon that has a representation of the chromat...

Страница 4: ...MultiChrom window and the SatView window 16 You should now see the 2000 40 MS acquisition window If necessary click on the Show Keypad button to see the acquisition controls Check the status indicators If they are both green the system is ready for the next sample 17 Rinse the injection syringe 25 times with methanol SHUT DOWN PROCEDURE 1 Minimize the 2000 44 acquisition control window 2 Maximize ...

Страница 5: ...n M C2H2N M C2H2N The adduct peaks are usually much lower in intensity than your parent MH peak For some types of samples the normal M ion is also produced by a charge transfer process M C2H3N M C2H3N For example acetonitrile CI works well with compounds containing nitrogen atoms producing a characteristic MH ion and a confirmatory M 13 adduct ion probably from the reaction with CH3CNCH2 Saturated...

Страница 6: ...s improved The disadvantage of using decreased target ion counts and maximum hold times is less sensitivity Since ion trap mass spectrometers are about ten times as sensitive as quadrupole mass analyzers even when operating with decreased source controls you can still have comparable sensitivity to a quadrupole mass spectrometer How do you know if auto ionization is a problem for your sample Compa...