UHV-24/UHV-24p Ionization Gauge
DRAFT
13
Operating Specifications
Table 1 Normal Operating Requirements
Item
Requirement
Filament Voltage
3 to 4.5 V (AC or DC) - Varies with pressure and emission
current
Filament Current
2.5 to 3.5 A - Varies with pressure and emission current
Element Voltages
Collector voltage: V
c
= 0
Filament voltage: V
fil
= 28 VDC
Grid voltage, V
g
= +180 VDC
Degas Power
Requirements (40 W)
Grid Voltage: +600 VDC max
Emission Current (total): 67 mA DC
Filament Voltage: 7 V
Measurement
Specifications
Sensitivity:
❑
25/Torr (UHV-24), nominal
❑
20/Torr (UHV-24p), nominal
Emission Current:
❑
4 mA for widest measurement range 5x10-
10
Torr to
1x10
-4
Torr
❑
10 mA for pressures lower than 5x10-10 Torr
0.1 mA
for pressure over 1x10
-4
Torr
Materials
Filament: Tungsten or thoria-coated Iridium
Filament Supports: SST
Collector - tungsten:
❑
0.007" diameter for UHV-24
❑
0.002" for UHV-24p
Grid: SST
Feedthrough: SST, Alumina, Nickel Alloy
Содержание UHV-24
Страница 2: ...D R A F T 1 6 1 4 Copyright 2014 Agilent Inc UHV 24 UHV 24p Ionization Gauge...
Страница 4: ...UHV 24 UHV 24p Ionization Gauge This page intentionally left blank D R A F T 1 6 1 4...
Страница 6: ...UHV 24 UHV 24p Ionization Gauge This page intentionally left blank D R A F T...
Страница 10: ...UHV 24 UHV 24p Ionization Gauge This page intentionally left blank...
Страница 20: ...UHV 24 UHV 24p Ionization Gauge 20 D R A F T Table 2 Gas Correction Factor Table...
Страница 21: ...UHV 24 UHV 24p Ionization Gauge D R A F T 21 Table B 0 Gas Correction Factor Table Continued...
Страница 22: ...UHV 24 UHV 24p Ionization Gauge This page intentionally left blank D R A F T...
Страница 26: ...UHV 24 UHV 24p Ionization Gauge This page intentionally left blank...
Страница 27: ......