13
U1731A/U1732A Quick Start Guide
Capacitance (Parallel Mode), Test Frequency = 100 Hz/120 Hz
Capacitance (Parallel Mode), Test Frequency = 1 kHz
Range
Maximum
Display
Accuracy
Specified Note
Capacitance
DF
10 mF
19.99 mF
*
* This reading can be extended up to 1999 MAX display with accuracy that is not specified.
3.0% + 5 (DF<0.1)
10.0% + 100/Cx + 5 (DF<0.1)
After short cal.
1000 µF
1999.9 µF
†
† This reading can be extended up to 19999 MAX display with accuracy that is not specified.
1.0% + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
After short cal.
200 µF
199.99 µF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
20 µF
19.999 µF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
2000 nF
1999.9 nF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
200 nF
199.99 nF
0.7% + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
After open cal.
20 nF
19.999 nF
1.0% + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
After open cal.
Range
Maximum
Display
Accuracy
Specified Note
Capacitance
DF
1 mF
1.999 mF
*
* This reading can be extended up to 1999 MAX display with accuracy that is not specified.
3.0% + 5 (DF<0.1)
10% + 100/Cx + 5 (DF<0.1)
After short cal.
200 µF
199.99 µF
1.0% + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
After short cal.
20 µF
19.999 µF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
2000 nF
1999.9 nF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
200 nF
199.99 nF
0.7% + 3 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
-
20 nF
19.999 nF
0.7% + 5 (DF<0.5)
0.7% + 100/Cx + 5 (DF<0.5)
After open cal.
2000 pF
1999.9 pF
1.0% + 5 (DF<0.1)
2.0% + 100/Cx + 5 (DF<0.1)
After open cal.