U1732A Electrical Specifications
Accuracy is expressed as ± (% of reading + number of least significant digits) at
23 °C ± 5 °C and <75% R.H.
Resistance (Parallel Mode), Test Frequency = 100 Hz/120 Hz
Resistance (Parallel Mode), Test Frequency = 1 kHz/10 kHz
Range
Maximum
Display
Accuracy
Specified Note
@ 100 Hz
@ 120 Hz
10 M
W
9.999 M
W
2.0% + 8
*
* This specification is based on the battery operation.
2.0% + 8
After open cal.
2000 k
W
1999.9 k
W
0.5% + 5
0.5% + 5
After open cal.
200 k
W
199.99 k
W
0.5% + 3
0.5% + 3
-
20 k
W
19.999 k
W
0.5% + 3
0.5% + 3
-
2000
W
1999.9
W
0.5% + 3
0.5% + 3
-
200
W
199.99
W
0.8% + 5
0.8% + 5
After short cal.
20
W
19.999
W
1.2% + 40
1.2% + 40
After short cal.
Range
Maximum
Display
Accuracy
Specified Note
@ 1 kHz
@ 10 kHz
10 M
W
9.999 M
W
2.0% + 8
*
* This specification is based on the battery operation.
3.5% + 10
After open cal.
2000 k
W
1999.9 k
W
0.5% + 5
2.0% + 10
After open cal.
200 k
W
199.99 k
W
0.5% + 3
1.5% + 5
-
20 k
W
19.999 k
W
0.5% + 3
1.5% + 5
-
2000
W
1999.9
W
0.5% + 3
1.5% + 5
-
200
W
199.99
W
0.8% + 5
2.0% + 10
After short cal.
20
W
19.999
W
1.2% + 40
2.5% + 200
After short cal.