17
Technical Information
Profibus
204/224
TV 301-P Navigator Controller User Manual
/ 87-901-001-01 (B)
Figure 20
Table 9
Internal link state machine
Status Meaning
Power on
Interface set-up
Src addr
Search for the controller serial address
Identify
Identify controller model and protocol
Data ex
Regular data exchange
The two machines continuously share I/O data with the relative bus,
and interact each other only in following conditions:
ProfiBus Data Exchange is suspended (static diagnosis activated)
if the internal serial link in not established
Internal serial link Data Exchange is suspended (only for output
data) if the ProfiBus State machine is not in Data Exchange mode
As a general rule, unrecoverable communication errors make each machine
to jump to the state immediately following the power-on state (WAIT_PRM or
SRC_ADDR)
Содержание TV 301-P
Страница 4: ...TV 301 P Navigator Controller 4 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 12: ...Contents 12 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 122: ...11 Felhasználói Kézikönyv Megsemmisítés 122 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 152: ...14 Návod na Obsluhu Likvidácia 152 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 162: ...15 Priročnik za Navodila Odlaganje opadkov 162 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 172: ...16 Instructions for Use Disposal 172 224 TV 301 P Navigator Controller User Manual 87 901 001 01 B ...
Страница 219: ......
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......