17 Technical Information
Connection J1 Remote I/O Interconnections
226/288
Turbo-V 2300 TwisTorr Rack Controller User Manual / 87-901-044-01
Input / Output Electrical Specifications
Figure 17 Input / Output electrical specifications
PURGE/VENT VALVES
: the Turbo-V 2300 TwisTorr Pumping
System can integrate the Purge and Vent valves. The two valves are
Normally Closed (N.C.) so if a power fail occurs, the valves will
remain closed.
The valves operating mode can be set by serial communication with
WIN 125 or using front panel interface (see following diagram for
details). The valve operating mode can be changed only with the
pump in STOP status.