Technical Information
Connection J1 Remote I/O Interconnections
17
Turbo-V 1001 Rack Controller User Manual / 87-901-033-01
215/288
Connection J1 Remote I/O Interconnections
Figure 7
Remote I/ O connector
All the input/output remote signals to/from the controller must be
connected at P1 mating connector. With the provided J1 mating
connector (shipped with pin 3 shorted with pin 9 and pin 4 shorted
with pin 15) make the connection with AWG 24 (0.25 mm
2
) or smaller
wire to the pins indicated in the figure to obtain the desired
capability. It is a 15-pins D type connector; the available signals are
detailed in the table, the following paragraphs describe the signal
characteristics and use.
Tab. 3
PIN N.
SIGNAL
INPUT/OUTPUT
1
START/STOP (+)
IN
2
START/STOP (-)
IN
3
INTERLOCK (+)
IN
4
INTERLOCK (-)
IN
5
SPEED SETTING (+)
IN
6
SPEED SETTING (-)
IN
7
SOFT START (+)
IN
8
SOFT START (-)
IN
9
+24Vdc
OUT
10
SET POINT (RELAY)
OUT
Содержание Turbo -V 1001