TriScroll 320 Series Vacuum Pump 

2

Installation Requirements

Do not install or use the pump in an environment
exposed to atmospheric agents (rain, snow, ice), dust,
aggressive gases, or in explosive environments or those
with a high fire risk. 

During operation, the following environmental
conditions must be respected:

Operation

In order to reach ultimate vacuum, the pump must be left
running for about an hour with the inlet sealed.

Unlike conventional oil-sealed pumps, Agilent’s dry scroll
pumps do not have fluid to cleanse them of accumulated
dust and debris. Run the pump periodically at atmosphere
for a minute or two to flush out the pump. Flush the pump
regularly and adjust this schedule according to your
specific conditions.

Start up Procedure

1. Ensure that the vacuum system isolation valve is 

closed.

2. Turn on power to the pump.

3. Open the isolation valve.

Shutdown Procedure

1. Close the vacuum system isolation valve. This 

prevents debris in pump from being transported into 
the vacuum system.

2. Turn off pump.

Maintenance

Personnel responsible for pump operation and
maintenance must be well-trained and aware of the
accident prevention rules.

If a pump is to be discarded, it must be disposed of in
accordance with specific national and local standards.

Temperature:

+5 °C to +40 °C (41 °F to 104 °F)

Relative humidity:

0 to 95% (non-condensing)

CAUTION

As supplied from the factory, the pump 
is configured for low voltage. Verify that 
the configuration matches the supply 
voltage. 

If voltage changeover is required, 
configure the voltage as described in 
“Electrical Connections”  on page 7.

WARNING

During operation the outer surface of 
the motor housing can become hot. 
Avoid touching the motor housing 
during pump operation.

WARNING

The pump is designed for operation 
with neutral or noncorrosive fluids. It 
is absolutely forbidden to use it with 
potentially explosive, inflammable or 
poisonous substances.

WARNING

Death may result from contact with 

high voltages. Always take extreme care 
and observe the accident prevention 
regulations in force.

When the machine is powered up, 

be careful of moving parts and high 
voltages. 

If you have to perform maintenance 

on the pump after a considerable time 
in operation, allow the pump to cool as 
the temperature of the outer surface 
may be in excess of 60 °C (140 °F).

Always disconnect your power 

supply to the pump before beginning 
maintenance work.

NOTE

Before returning the pump to the factory 
for repair, the “Health and Safety” sheet 
attached to this instruction manual must 
be completed and sent to the local sales 
office. A copy of the sheet must be 
inserted in the pump package before 
shipping.

Содержание TriScroll 320 Series

Страница 1: ...INSTALLATION AND OPERATION MANUAL Vacuum Products Division TriScrollTM 320 Series Dry Scroll Vacuum Pump Manual No 699904342 Revision F October 2014...

Страница 2: ...Copyright 2014 Agilent Inc TriScroll 320 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent Inc...

Страница 3: ...earing Purge 9 Purge Kit 9 Troubleshooting 10 Maintenance 11 General Information 11 Related TriScroll Manuals 11 Maintenance and Tooling Kits 11 Factory Service Options 12 Accessories 12 Contacting Ag...

Страница 4: ...TriScroll 320 Series Vacuum Pump This page intentionally left blank...

Страница 5: ...ocol Storage When transporting and storing the pump the following environmental requirements should not be exceeded Preparation for Installation The pump is supplied in a special protective packing If...

Страница 6: ...ance with specific national and local standards Temperature 5 C to 40 C 41 F to 104 F Relative humidity 0 to 95 non condensing CAUTION As supplied from the factory the pump is configured for low volta...

Страница 7: ...n Female 1 4 National Pipe Thread NW16 adapter provided Gas ballast Female 1 4 National Pipe Thread 40 Micron sintered filter provided Ambient operating temperature 5 C to 40 C 41 F to 104 F Storage t...

Страница 8: ...TriScroll 320 Series Vacuum Pump 4 Figure 1 Interface Drawing with Dimensions Figure 2 Pumping Speed Curves...

Страница 9: ...13 Motor Electrical Cover Unpacking and Inspection The shipping container is a double carton 1 After opening the outer box remove the foam packing 2 Slit open the inner box 3 Lift the pump with the p...

Страница 10: ...er capacity The circuit breaker must be suitably located and easily reached The circuit breaker must be marked at the disconnecting device for the equipment with the on off symbol I O and meet the rel...

Страница 11: ...ct the motor to match your supply voltage Two options are available to strain relieve the electrical supply cable The cable can be mechanically held under the electrical cover in the groove provided T...

Страница 12: ...the vacuum chamber Use a bellows to provide both vibration isolation and strain relief between the pump and the vacuum chamber Pump Exhaust A female 1 4 National Pipe Thread exhaust fitting is located...

Страница 13: ...Scroll motor s inter nal thermal switch to actuate the VPI Valve Gas Ballast The gas ballast port item 9 on Figure 3 on page 5 is sealed on the TriScroll 320 For applications where a lot of water is b...

Страница 14: ...power supply cable Defective motor Inspect Contact Agilent Poor ultimate pressure System leak Locate and repair leak Water in pump Flush pump with air or dry nitrogen Gas ballast plugged Replace breat...

Страница 15: ...maintenance on TriScroll pumps is provided in kit form A description of each kit and ordering information is provided in Table 6 Table 5 Other Related Manuals Title Applicable TriScroll Model Part Num...

Страница 16: ...anual Contacting Agilent In the United States you can contact Agilent Customer Service at 1 800 882 7426 See the back cover of this manual for a listing of our sales and service offices Visit our web...

Страница 17: ...re product e g inlet screens vent valves x Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue x If ordering an Advance Exchange product p pleas...

Страница 18: ...or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled ou...

Страница 19: ...ther Vacuum leak Other Section 6 ADDITIONAL TERMS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Produc...

Страница 20: ...TriScroll 320 Series Vacuum Pump This page intentionally left blank...

Страница 21: ......

Страница 22: ...n 54 10040 Leini Torino ITALY Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Japan Agilent Technologies Japan Ltd Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4 1...

Отзывы: