Window Protocol
Description
Communication Format
8 data bit
no parity
1 stop bit
baud rate: 600/1200/2400/4800/9600 programmable
Communication Protocol
The communication protocol is a MASTER/SLAVE type where:
Host = MASTER
Controller = SLAVE
The communication is performed in the following way:
1
the host (MASTER) send a M CRC to the controller (SLAVE);
2
the controller answer with an CRC to the host.
The MESSAGE is a string with the following format:
<STX>+<ADDR>+<WIN>+<COM>+<DATA>+<ETX>+<CRC>
Where:
NOTE
When a data is indicated between two quotes (‘...’) it means that the indicated data
is the corresponding ASCII character.