Technical Information
Interconnections
17
TPS-compact User Manual / 87-901-010-01 (F
)
281/312
Figure 30
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
§
reference quantity: frequency
§
thereshold: 867 Hz
§
hysteresis: 2 %
§
activation type: high level
§
delay time: 0 second
NOTE
The Navigator Software (optional) allows the operator to set all the
programmable feature.
How to Connect the Open Collector Input of the
Controller
Here below there are the typical connections of the open collector
input of TPS-compact to an external system. Two cases are
considered:
Содержание TPS-compact X3580-64000
Страница 3: ...TPS compact TPS compact User Manual 87 901 010 01 F 3 312 TPS compact...
Страница 4: ...TPS compact 4 312 TPS compact User Manual 87 901 010 01 F...
Страница 42: ...2 Gebrauchsanleitung Entsorgung 42 312 TPS compact User Manual 87 901 010 01 F...
Страница 58: ...3 Mode d emploi Mise au rebut 58 312 TPS compact User Manual 87 901 010 01 F...
Страница 74: ...4 Manual de istrucciones Eliminaci n 74 312 TPS compact User Manual 87 901 010 01 F...
Страница 90: ...5 Manual de Istru es Elimina o 90 312 TPS compact User Manual 87 901 010 01 F...
Страница 106: ...6 Bedrijfshandleiding Afvalverwerking 106 312 TPS compact User Manual 87 901 010 01 F...
Страница 164: ...10 Ohjek sikirja H vitt minen 164 312 TPS compact User Manual 87 901 010 01 F...
Страница 180: ...11 Felhaszn l i K zik nyv Megsemmis t s 180 312 TPS compact User Manual 87 901 010 01 F...
Страница 224: ...14 N vod na Obsluhu Likvid cia 224 312 TPS compact User Manual 87 901 010 01 F...
Страница 307: ......
Страница 308: ......
Страница 309: ......
Страница 310: ......
Страница 311: ......