Step 5 

After dispensing the injection solutions into the ports, with-

draw the tips from the ports carefully, holding the port loading 

guide throughout the procedure. Remove the A/D port loading 

guide and set aside for loading port D (if needed). Do not 

tap any portion of the XF Sensor Cartridge in an attempt to 

remove air bubbles from an injection port. This can cause 

unintended leaking of injection solutions from the injection 

ports.

NOTE: 

Forcefully dispensing injection solutions can over-

whelm surface tension at the injection port opening and result 

in leaking of  the solution from the injection port.

Step 6

Switch to the B/C port loading guide (red lettering). Orient 

with the letter ‘B’ (green arrow, Figure 7) in the upper left-hand 

corner, and place on top of the XF Sensor Cartridge. Repeat 

loading procedure outlined in steps 3-5 for ‘B’, and injection 

ports. If your experimental design requires an injection from 

Port D, remove the B/C port loading guide and place the A/D 

port loading guide on the XF Sensor Cartridge. Repeat loading 

procedure outlined in steps 3-5 for injection port D.. Remove 

and discard loading guide(s) once the entire XF sensor car-

tridge has been loaded.

Step 7

Visually inspect the injection ports for even loading (Figure 

8). The solution should be in the port; make sure there are 

no residual drops on top of the XF Sensor Cartridge. Once all 

injection solutions have been loaded according to your ex-

perimental design, carefully transfer the XF Sensor Cartridge 

(together with the XF Utility Plate) to the Seahorse XFe96/

XF96 Analyzer to start your assay.

IMPORTANT:

 Remove all port loading guides and lids for the 

XF Sensor Cartridge and XF Utility Plate before placement 

into the Analyzer.

Figure 7

 Orienting the B/C port loading guide.

1

A

B

C

E

G

B

D

F

H

3

5

7

9

11

2

4

6

8

10

12

1

A

C

C

E

G

B

D

F

H

3

5

7

9

11

2

4

6

8

10

12

Figure 8

 Visual verification of injection port loading.