background image

3

Step 3 

Place the A/D port loading guide (blue lettering) flat on top 

of the sensor cartridge. Orient the port loading guide so the 

letter ‘A’ (Figure 3) is in the upper left-hand corner. Use your 

fingertips to hold the outside edges of the port loading guide 

to stabilize during loading and to prevent pipette tips from 

dislodging the port loading guide (Figure 5).

Figure 5

 Dispensing Injection Solutions using the port loading guide.

Figure 6. Injection Port Mechanism and Proper Port Loading Techniques-  (a)

 Cross section of a single XF injection port. Note the open orifice at the bottom of 

the port. Injection solutions are retained in the ports via surface tension. 

(b)

 Pressurized air is used to force the solution from the port during an injection. 

(c) 

Do 

NOT insert pipettes tips fully into the ports if not using a loading guide. 

(d)

 If using the XF Port Loading Guide, place the pipette tips vertically into the Port Loading 

Guide holes as shown. 

(e)

 If not using the XF Port Loading Guide, then allow the pipette tips to gently touch the inside of the ports, resting the side of the pipette 

tips on the port rims. 

(f, g)

 Acceptable positions of injection solution in port after loading. 

(h)

 Unacceptable position of injection solution in port after loading. 

Figure 4

 XFe96 Sensor Cartridge Port designations.

If you are not using a port loading guide, port A is the upper-

left port in each well quadrant (Figure 4). 

Step 4

Using a multi-channel pipette, first ensure tips are securely 

fitted onto the pipette before pipetting injection solutions. 

Position the pipette tips (filled with your injection solutions) 

into the desired column in the port loading guide, and orient 

the tips at a very slight angle less than 5° (Figures 5 and 6d). 

Insert the tips as far as they will go without resistance into the 

holes and dispense the injection solutions. Do not force the 

tips completely into the holes of the port loading guide  

(Figure 6c).

Seahorse

A

C

E

G

B

D

F

H

A

B

C

D

Port designation

for each well

e

f

g

h

a

Injection

Port

Without

Loading

Guide

OR

d

With

Loading

Guide

Injection

Manifold

Injection

Port

Orifice

b

c

Figure 3

 Orienting the A/D port loading guide.

1

A

A

C

E

G

B

D

F

H

3

5

7

9

11

2

4

6

8

10

12

1

A

D

C

E

G

B

D

F

H

3

5

7

9

11

2

4

6

8

10

12

When not using a port loading guide, position the pipette tips 

as shown in Figure 6e. Do not fully insert the pipet tips into 

the port as this may result in accidentally forcing injection 

solutions through the port orifice (Figure 6c).

Отзывы: